7 Maintenance
K-375 / K-376 / K-377
Operation Manual, Version E
139
NOTE
Make sure that one of the clips of the
holder presses on the spring inside the
sampler arm. (See arrow in the figure
beneath.)
Otherwise the sample tube detection will
not work!
7.6 Replace every two years
7 6 1
Replacement of the transfer connection
!
WARNING
Serious chemical burns by corrosives. Risk of burns by hot steam.
∙ Never operate the K-375 together with a sampler while the sample transfer and/
or steam hose is getting porous or cracked.
!
WARNING
Serious chemical burns by corrosives.
∙ Always wear protective goggles.
∙ Always wear protective gloves.
∙ Always wear protective clothes.
The transfer connection consisting of the transfer hose, the steam hose and the black protective
sleeve shall be replaced at least every two years or if required.
To replace the transfer connection proceed as follows: