6 Operation
K-375 / K-376 / K-377
Operation Manual, Version E
96
Within the settings for the blank correction,
∙ switch “Blank correction”
ON
∙ switch “Auto blank generation”
OFF
∙ select
NO
for “Use last measured blank”
∙ confirm your settings with
OK
.
Press
MANUAL
Enter a name for the manual blank value.
Enter the volume for the blank in [mL]
Confirm your settings with
OK
.
The entered blank value is now automatically selected and shown in the list of blanks in the Blank
Correction screen.
Manually entered blanks are listed in the blank list with type “manual”.
NOTE
If you select
Yes
for
“
Use last measured blank“ the next measured value for a sample of Type
“
Blank“
will be used for the correction of all subsequent sample determinations. All sample determinations up
to the next blank determination will still be corrected using the currently selected blank value.
Defining Mean blanks
Mean blank values can be calculated over two or more measured blank values. To define a Mean
blank value, proceed as follows:
Enter the
Blank Correction
screen.
Press
SETTINGS
.