5 Putting into operation
K-375 / K-376 / K-377
Operation Manual, Version E
45
∙ Connect the power cable to the power connection
1
.
∙ Connect the level sensors to the corresponding connectors
7
.
NOTE
Unlike the level sensors for the storage tanks of H
2
0, NaOH and H
3
BO
3
, the presence of the level
sensors for the waste containers has to be indicated within the software! (See section “Peripherals”
in chapter 6.9.1)
∙ Connect the dosing unit for the acid to connector
9
.
∙ Connect the RS232 cable to the sampler (if present) to the corresponding connector
2
.
∙ Connect any additional peripherals
according to the description in figure 5.1.
5 2 2
Connections of the K-376 / K-377
(left rear side of the housing)
(right rear side of the housing)
1
Power switch K-376 / K-377
2
Power connection K-376 / K-377
3
Fuses (2 x 3A)
4
RS232 connection to K-375
5
Toggle switch (see chapter 8.3)
Fig. 5.2: Electrical connections of the K-376 / K-377
NOTICE
Risk of device damage by wrong voltage.
∙ Make sure that the voltage on the socket corresponds to the voltage given on
the type plate of the instrument.
∙ Always connect the device to an earthed socket. External connections and
extension cables must be provided with an earthed conductor lead (3-pole
couplings, cable or plug equipment) as the mains lead has a molded plug, thus
avoiding risks due to inadvertent defective wiring.
∙ Make sure that no electric sparks form in the device or its surroundings as they
might damage the instrument.
On the KjelSampler K-376 / K-377
∙ Connect the power cable to the power connection
2
∙ Connect the cable to the K-375 device to the RS232 connector
4