6 Operation
K-375 / K-376 / K-377
Operation Manual, Version E
113
Peripherals
Select
Yes
for all present peripherals:
∙ Level sensor for receiver waste container
∙ Level sensor for sample tube waste container
∙ Sampler
∙ Balance
∙ Chiller
∙ Printer
NOTE
Unlike the level sensors for the storage tanks of
H
2
0, NaOH and H
3
BO
3
, the level sensors for the
waste containers have to be activated!
If a chiller is connected to the K-375 it is strictly
necessary to select “Yes” for “Chiller used” in
order to have the cooling water valve permanently
opened.
If a sampler is connected to the instrument, select
the type (1 tray for K-376, 2 trays for the K-377)
and enable or disable the auto cleaning functionality
of the sampler after each sample measurement.
If a balance is connected to the instrument, specify
the communication settings for the balance.
∙ the Baud rate
∙ the number of Data bits (7 or 8)
∙ the Parity (none, even or odd)
∙ the number of Stop bits (1, 1.5 or 2)
NOTE
For details about the single parameters, please
refer to the operation manual of the balance.