5 Putting into operation
K-375 / K-376 / K-377
Operation Manual, Version E
46
5.3 Transfer connection K-376 / K-377 to K-375
The transfer connection between the K-375 and the K-376 or K-377 sampler consists of two hoses,
the white transfer hose and the red steam hose.
Both hoses have to be connected to the K-375 as well as to the sampler (K-376 or K-377) and
secured with hose clamps. The K-376 is delivered with both hoses premounted to the device.
!
WARNING
Serious chemical burns by corrosives. Risk of burns by hot steam.
∙ Never operate the K-375 together with a sampler while the sample transfer and
steam hoses are missing, defective, or incorrectly mounted.
∙ Make sure there is always enough room for a free movement of the sampler
arm – if the sampler arm collides with any object during movement, the transfer
hose and/or the steam hose may break!
5 3 1
Connecting the K-376 to the K-375
∙ Fix the transfer hose support
1
with the
screw on the valve
2
on the rear side of the
K-375.
∙ Guide both hoses
3
and
4
through the
transfer hose support.