5 Putting into operation
K-375 / K-376 / K-377
Operation Manual, Version E
53
5.6 Positioning of the dosing tip
Mount the spacer onto the titrant dosing tip to adjust the positioning of the outlet and place it in the
receiving vessel. It should be positioned in the same height as the stirrer.
Fig. 5.12: Mounting of the spacer at the dosing tip
NOTE
The dosing tip must not touch the bottom of the receiving vessel, as this would block the outlet.
5.7 Storage tank connection
To connect the storage tanks, proceed as follows:
∙ Cut the Nylflex tube into pieces to the appropriate length.
∙ Insert a PTFE suction tube into the Nylflex tube.
∙ Push a EPDM sealing ring over the Nylflex tube.
∙ Now fasten the tubes to the tank with the red screw cover.
The storage tanks should not be positioned higher than the device itself and not lower than
1 meter below the instrument.
Fig. 5.13: Tank connection
All pumps are self-priming, no overpressure is necessary at the tanks.