8 Troubleshooting
K-375 / K-376 / K-377
Operation Manual, Version E
150
10’208
Sensor ‘steam pressure’ out of order
Fault in steam pressure measurement,
AD-converter or other hardware component.
Call service.
10’217
Analog digital
converter out of order
Fault in AD-converter or other hardware
component. Call service.
10’300
No cooling water flow detected.
Please turn on water tap.
Check cooling water supply. Turn on tap or
chiller.
10‘301
Aspiration error: No vacuum detected Check the system for leaks
10‘302
Cooling water flow too low
Assure higher flow rate or check parameters in
Settings/Peripherals/Cooling water settings
10‘303
Low pressure during distilling
System pressure is below 150 mbar. Check
for leaks or call service.
10‘311
Pump H2O has no current
Faulty water pump. Exchange pump or call
service.
10’312
Pump NaOH has no current
Faulty NaOH pump. Exchange pump or call
service.
10’314
Pump H3BO3 has no current
Faulty boric acid pump. Exchange pump or
call service.
12’001
Valve steam (Y1) out of order
Faulty valve or wire harness. Call service.
12’002
Valve cooling water in (Y5) out of
order
Faulty valve or wire harness. Call service.
12’003
Valve sampler steam (Y6) out of order Faulty valve or wire harness. Call service.
12’004
Valve sampler transfer (Y7) out of
order
Faulty valve or wire harness. Call service.
12’005
Valve 5 (not used) out of order
Faulty valve or wire harness. Call service.
12’006
Valve waste sample tube (Y2) out of
order
Faulty valve or wire harness. Call service.
12’007
Valve aspiration in (Y3) out of order
Faulty valve or wire harness. Call service.
12’008
Valve receiver (Y4) out of order
Faulty valve or wire harness. Call service.
12’009
Valve H2O injection (Y8) out of order
Faulty valve or wire harness. Call service.
12’010
Valve H2O sample tube (Y9) out of
order
Faulty valve or wire harness. Call service.
12’011
Valve waste receiver (Y10) out of
order
Faulty valve or wire harness. Call service.
13’001
27V power supply overcurrent
Faulty electronic board. Call service.
13’002
Ventilator power overcurrent
Fan short circuit. Call service.
13’003
Ventilator electronics blocked
Check for blockage or call service
13’004
Ventilator interior blocked
Check for blockage or call service