6 Operation
K-375 / K-376 / K-377
Operation Manual, Version E
104
Parameters for Boric acid titration with
colorimetric sensor
Select
Boric acid titration
for the parameter
“Titration type“. The volume of the boric acid to be
dosed into the receiving vessel can be specified
and the
titration solution
can be selected from
the tritration solutions list (see chapter ”6.8.2 Volu-
metric Solutions”).
Select
Colorimetric
for the parameter “Sensor
type“.
For the “Titration mode” either Online or Standard
can be selected:
Online:
The titration is already started, while the
distillation is running. With the parameter “Titration
start time“ a delay time for the titration start can
be defined. It is recommended to set a delay of 90
seconds.
Standard:
The titration is started sequential after
the distillation is finished.
Furthermore the stirrer speed for the titration and a
titration start volume can be specified. The speci-
fied titration start volume is dosed into the receiving
vessel at the beginning of the titration when
samples are analyzed
With the parameter “Titration algorithm” one of two
available algorithms for the titration can be selected:
Optimal:
This algorithm has been optimized with
respect to the duration of the titration and the accu-
racy of the result. It is recommended for samples
with a higher nitrogen content.
Normal:
This algorithm has been optimized with
respect to the accuracy of the result but will require
a longer time than the optimal algorithm. The
normal algorithm is recommended for samples with
low nitrogen content and when using titration solu-
tions with high concentration (e.g. 0.5 N).