6 Operation
K-375 / K-376 / K-377
Operation Manual, Version E
106
Determination Parameters
With the determination parameters the Determination mode (either “Standard” or “Direct distillation”)
can be determined.
In addition two sets of result units (each set consisting of one mass- and one volume-related unit) can
be selected for the standard determination mode.
Click on “Determination mode“ and select either
“Standard“ for a determination with a previous
digestion or “Direct distillation“ for a determination
without previous digestion.
For a direct distillation an additional factor and
regression factor can be specified for result calcula-
tion.
NOTE
The result for a direct distillation is calculated
following the linear equation
y= a·x+b
where
a
is
given by the direct distillation factor and
b
by the
regression factor.
The four result units (mass and volume) for the
determination mode “Standard” can be adapted
by clicking on it and selecting a corresponding
unit from the result units list. (See chapter “6.9.1
Settings“)