6 Operation
K-375 / K-376 / K-377
Operation Manual, Version E
95
6 7 3
Blank Correction
The blank correction can be switched on and off:
HOME
Blank Correction
SETTINGS
ON
Blank correction of results is switched on.
OFF
Blank correction of results is switched off –
no blank correction at all will be performed.
Within the Blank Correction main screen a list of the latest blanks can be viewed. By adapting the
parameter “Blanks in list” under
SETTINGS
it can be determined how many blanks will be shown in
this list:
-
Will decrease the number of displayed blanks
by 10.
+
Will increase the number of displayed blanks
by 10.
A maximum number of 90 blanks can be listed.
In general four different possibilities for the determination of the active blank value for the automatic
correction of results are available:
∙ Blank values can be measured (type: measured).
∙ Blank values can be entered manually (type: manual).
∙ Blank values can be calculated as mean value of freely selectable measured blanks (type:
mean).
∙ Blank values can be determined automatically by the system (type: automatic).
The type and the value of the blank currently used for result correction is always displayed in the
results section of the status view:
SHOW STATUS
RESULTS
(See section 6.5.1)
Entering Manual blanks
To enter a blank value manually (e.g. for a blank value that has not been determined with the instru-
ment), proceed as follows:
Enter the
Blank Correction
screen.
Press
SETTINGS
.