6 Operation
K-375 / K-376 / K-377
Operation Manual, Version E
118
Demo mode
Select
On
to switch to the Demo mode and
Off
to
leave the Demo mode.
NOTE
In demo mode no chemicals are used or dosed
and determinations are only simulated.
The demo mode is automatically deactivated as
soon as the instrument is switched off.
6 9 3 Diagnostics
In the system diagnostics the actual status of all
system components like valves, pumps, sensors
and switches or fans can be checked.
Each item is displayed in a list and, if currently
active, a green dot is displayed in front of it.
With the button
SERVICE MODE
one can switch
to the service mode to actively set each system
component into operation and check it for proper
functioning.
NOTE
The diagnostics dialogs are only accessible for user
with administrator rights! (For users with operator
rights the diagnostics area is “read-only”).
CAUTION
While the service mode is activated, some security-relevant functions are stopped. Executing single
test functions is therefore subject to the operator‘s responsibility. The service mode will be switched
off, as soon as the diagnostics menu is left.