7 Maintenance
K-375 / K-376 / K-377
Operation Manual, Version E
129
7 3 2
Checking the distillate amount
∙ Carry out preheating (three times) so that the system is warm before performing this test.
∙ Create a new method using the following parameters:
H
2
O volume:
0 mL
NaOH volume:
0 mL
Reaction time:
0 s
Distillation mode:
Fixed time
Distillation time:
300 s
Stirrer speed distillation:
5
Steam output:
100 %
Titration type:
None
Aspiration sample tube:
Yes
Aspiration receiving
vessel:
No
∙ Run the method with an empty sample tube and empty receiving vessel.
∙ Measure the distilled quantity in the receiving vessel using a graduated cylinder.
NOTE
The distillate amount with above parameters must be
≥
130 mL.