6 Operation
K-375 / K-376 / K-377
Operation Manual, Version E
115
Import and export
Specify the target for the manual or automatic data
storage. You may either select a USB device or a
network directory or both in parallel:
∙ Select
Yes
for „Use USB device“ for the data
storage to a USB device.
∙ Select
Yes
for „Use network data share“ for
data storage on a network place.
∙ For an automatic data export after each
sample determination select
Yes
for „Export
results automatically“.
With „Test Export“ the validity of the settings can be
checked.
NOTE
For the network settings configuration the help of
an IT specialist of the company may be required!
Results are exported from the K-375 using xml-
format. Most LIM systems are able to handle this
format. If more details regarding the data format
are required refer to the document “Manual – Data
export K-375”, which can be obtained from any
authorized BUCHI representative.
User administration
Create, delete or rename users. Assign user rights
to each user.
NOTE
The first created user has to be of type “Adminis-
trator”. After different users have been configured,
the “User management” is active and the Logout
functionality becomes visible on the home screen.
As soon as a user account has been created the
system automatically creates a default “BUCHI
Administrator” account. This account can be used
in case all users with own accounts forgot their
passwords. The password for this default adminis-
trator account (a four digit code which changes on
a daily base) can be obtained from any authorized
BUCHI representative.