6 Operation
K-375 / K-376 / K-377
Operation Manual, Version E
91
Using the
BALANCE
button the weight of samples
can be taken over from a connected balance:
∙ Select all samples
∙ Press
BALANCE
– all blanks and control
blanks are automatically deselected (no weight
is needed for blanks)
∙ Place the first sample on the balance and
press
Enter
on the balance – the first weight
is taken over from the balance and entered
into the first checked sample in the list.
∙ Proceed with the next sample
∙ When all sample weights have been taken
over, the balance mode is automatically left.
If certain samples of a rack that is already in prog-
ress have to be determined immediately this can be
done using the
Split Rack
functionality:
Press
PAUSE
to stop the determination of the
sequence.
Select the samples to be determined immediately
and press
SPLIT RACK
.
The selected samples will be deleted from the
sequence and will be inserted into the rack-step of
a new created sequence at the same position of the
rack.
The newly created Split-Sequence can be started
to determine express samples – afterwards the
previous sequence can be continued.
NOTE
Using the button
EDIT MODE
it is possible to edit
samples of a rack that is not yet being processed
during a running sequence.