table 21: flow error description and recommended actions
Error Code
Problem
Cause
Recommended Action
E23: Multi Channel Accuracy
Two or more measurement
channels are in error
Two or more measurement
channels are in error;
accuracy of the
measurement may be
compromised because the
meter is using a sister chord
substitution
Check individual channel
errors, refer to this table for
recommended actions to
correct channel errors
E27: Invalid K-Table
K-Table is invalid
The entered K-table is invalid
Check the K-table values
and ensure the Velocity or
Reynolds Number in the
table is in ascending order
E28: Software Fault
Software malfunction
This is a Software
malfunction.
This condition is not
self-recovering and will not
automatically correct itself.
Try power cycling the meter.
If error persists after power
cycle, contact Panametrics
factory.
E29: Velocity Warning
The measured velocity
exceeds programmed
warning limits
This error may be caused
by incorrect programming,
poor flow conditions and/or
excessive turbulence
Make sure the actual
flow rate is within the
programmed Warning
limits (Refer Programming
Chapter). Refer to “Fluid
and Pipe Problems” and
“Transducer Problems”
sections to correct any
issues
E31: Not Calibrated
The flow meter has not been
calibrated
The flow meter has not been
calibrated at the factory
and hence not making
measurements. Please
contact Panametrics factory
The condition is not self-
recovering and will not
automatically correct itself.
Contact Panametrics factory
to get more information on
the meter setup
57
Содержание Panametrics PanaFlow LC
Страница 1: ...PanaFlow LC User s manual 910 327 Rev A...
Страница 2: ...ii...
Страница 3: ...PanaFlow LC Panametrics liquid ultrasonic flowmeter User s manual 910 327 Rev A Sep 2019 iii...
Страница 4: ...no content intended for this page iv...
Страница 12: ...no content intended for this page xii...
Страница 44: ...Figure 58 Setting up pulse output 32...
Страница 49: ...Figure 60 Setting up alarm output 37...
Страница 80: ...no content intended for this page 68...
Страница 82: ...no content intended for this page 70...
Страница 86: ...no content intended for this page 74...
Страница 98: ...Table 30 Service Record Date Description of Service Performed By 86...
Страница 104: ...no content intended for this page 92...
Страница 106: ...no content intended for this page 94...
Страница 114: ...no content intended for this page 102...
Страница 138: ...no content intended for this page 126...
Страница 140: ...no content intended for this page 128...
Страница 145: ...no content intended for this page 133...