Attention!
Before entering a password, make sure the
XMIT_TB transducer block is in active mode.
To Enter configuration mode, complete the following steps:
1. Select the
Resource Block > Others
tab.
2. Select
SYSTEM_OPT_TYPE
and set to
Option FI
.
3. Select
SYSTEM_REQ_LEVEL
and set to
Admin
or
Operator
.
4. Enter the Admin or Operator password into the
SYSTEM_
REQ_PASS
field.
5. Select
Cancel
from the
SYSTEM_REQ_CMD
drop down
box.
6. Select
Login
from the
SYSTEM_REQ_CMD
drop down
box.
7. Click on the
Write Changes
button.
8. Verify that
S1:In Config Mode
appears on XMT1000
display. You should now be able to edit fields with
Admin
privileges.
To Edit the fields in the Transducer Blocks, complete the
following steps:
1. Select or enter the
new value
.
2. Click on the
Write Changes
button.
3. Return to the
Resource Block > Others
tab and select
Commit from the
SYTEM_REQ_COM
drop down box
4. Click on the
Write Changes
button.
To Exit configuration mode, complete the following steps:
1. Select Cancel from the
SYSTEM_REQ_CMD
drop down
box.
2. Click on the
Write Changes
button.
Note:
The XMT1000 will automatically exit configuration
mode after 5 minutes of inactivity.
H.4.3 NAMUR NE107
The
NAMUR NE107
recommendation specifies that detailed
device-specific diagnostics are summarized as four simple
status signals. The diagnostics are set to defaults by
Panametrics, but they can be modified to any other level by
the user. The four status signals are:
•
Failed:
This category is typically used for hardware or
software failures. The meter output is not valid. Consult
with the factory for a resolution.
•
Offspec:
This category is typically used for application,
installation, or process problems. Consult the
troubleshooting section of this appendix or contact
Panametrics customer service for assistance.
•
Check:
This category means the output of the device
is invalid due to on-going work on the device, such as
programming, etc.
•
Maintenance:
This category is typically used to assign
parameters that are in good status but which may fall
out of specification due to some process condition or
wear factor. There are no diagnostics set as default in
this category.
The status signals (see Figure 133 below) can either be
reported as errors (
ACTIVE
) or masked when the error occurs
(
MASK
). There are no default settings for the
MASK
bits.
Figure 133: NAMUR NE107 configuration in resource block
109
Содержание Panametrics PanaFlow LC
Страница 1: ...PanaFlow LC User s manual 910 327 Rev A...
Страница 2: ...ii...
Страница 3: ...PanaFlow LC Panametrics liquid ultrasonic flowmeter User s manual 910 327 Rev A Sep 2019 iii...
Страница 4: ...no content intended for this page iv...
Страница 12: ...no content intended for this page xii...
Страница 44: ...Figure 58 Setting up pulse output 32...
Страница 49: ...Figure 60 Setting up alarm output 37...
Страница 80: ...no content intended for this page 68...
Страница 82: ...no content intended for this page 70...
Страница 86: ...no content intended for this page 74...
Страница 98: ...Table 30 Service Record Date Description of Service Performed By 86...
Страница 104: ...no content intended for this page 92...
Страница 106: ...no content intended for this page 94...
Страница 114: ...no content intended for this page 102...
Страница 138: ...no content intended for this page 126...
Страница 140: ...no content intended for this page 128...
Страница 145: ...no content intended for this page 133...