2.11.3.1 Transducer wiring (ATEX installations)
1. Thread ends of armored cable into the meter head
and the other into the junction box on the back of the
transducer (see Figure 1 on page 1 for reference).
2. Terminate connections of the transducer end of each
cable to each of your clamp on transducers.
3. At the meter end of the cable secure the red wire of
channel one downstream RG62 cable to the CH1 (red
box in Figure 41) DN terminal of the XMT1000 backplane.
4. Secure Black wire of channel one downstream RG62
cable to the CH1 (red box in Figure 41) RTN terminal
directly below the DN terminal of the XMT1000
backplane.
5. Secure red wire of channel one upstream RG62 cable to
the CH1 (red box in Figure 41) UP terminal of the XMT1000
backplane.
6. Secure Black wire of channel one upstream RG62 cable
to the CH1 (red box in Figure 41) RTN terminal directly
above the UP terminal of the XMT1000 backplane.
7. At the meter end of the cable secure the red wire of
channel two downstream RG62 cable to the CH2 (blue
box in Figure 41) DN terminal of the XMT1000 backplane.
8. Secure Black wire of channel two downstream RG62
cable to the CH2 (blue box in Figure 41) RTN terminal
directly below the DN terminal of the XMT1000
backplane.
9. Secure red wire of channel two upstream RG62 cable
to the CH2 (blue box in Figure 41) UP terminal of the
XMT1000 backplane.
10. Secure Black wire of channel two upstream RG62 cable
to the CH2 (blue box in Figure 41) RTN terminal directly
above the UP terminal of the XMT1000 backplane.
CH2
CH1
Figure 43: PanaFlow LC2 channel set up
2.11.4 Three channel set up
Components of a three channel PanaFlow LC will be: a meter
head, 6 conductor “trunk” cable, 3 channel junction box, 6
transducer cables and 3 sets of transducers and clamping
fixtures. Mount meter head and junction box to a two inch
post using included hardware or another location following
local electrical and building codes.
Install transducers as mentioned in Section 2.7 through
Section 2.9. Prepare to wire transducer cables.
Note:
US/ Canada installations will be required to provide
conduit per their particular location. European installations
will include armored cable.
2.11.4.1 Transducer wiring (ATEX installations)
1. Thread MCX end of 6 conductor cable trunk into
XMT1000 meter head and the BNC connector end of the
cable into the 3 channel Junction box as seen in Figure
44.
XMT1000 Meter Head
3 Channel Junction Box
Trunk Cable
Figure 44: XMT1000 Trunk cable placement
17
Содержание Panametrics PanaFlow LC
Страница 1: ...PanaFlow LC User s manual 910 327 Rev A...
Страница 2: ...ii...
Страница 3: ...PanaFlow LC Panametrics liquid ultrasonic flowmeter User s manual 910 327 Rev A Sep 2019 iii...
Страница 4: ...no content intended for this page iv...
Страница 12: ...no content intended for this page xii...
Страница 44: ...Figure 58 Setting up pulse output 32...
Страница 49: ...Figure 60 Setting up alarm output 37...
Страница 80: ...no content intended for this page 68...
Страница 82: ...no content intended for this page 70...
Страница 86: ...no content intended for this page 74...
Страница 98: ...Table 30 Service Record Date Description of Service Performed By 86...
Страница 104: ...no content intended for this page 92...
Страница 106: ...no content intended for this page 94...
Страница 114: ...no content intended for this page 102...
Страница 138: ...no content intended for this page 126...
Страница 140: ...no content intended for this page 128...
Страница 145: ...no content intended for this page 133...