4.10 Option I/O errors
Table 27: Option I/O errors description
Error Code
Error Message
Description
A1:AnalogCh(S2:3) Error!
ADC Channel(S2:3) is not responding
Analog input /RTD input is not working.
If error persists after power cycle,
contact Panametrics factory
A2:AnalogCh (S2:4) Error!
ADC Channel(S2:4) is not responding
Analog input /RTD is not working. If error
persists after power cycle, contact
Panametrics factory
A3:AnalogCh (S2:1) Error!
DAQ Channel (S2:1) is not responding
Analog output (4-20mA) is not working.
If error persists after power cycle,
contact Panametrics factory
A4:AnalogCh (S2:2) Error!
DAQ Channel (S2:2) is not responding
Analog output (4-20mA) is not working.
If error persists after power cycle,
contact Panametrics factory
A6:(S2:3)Ch Not Calibrated
Error occurs when Analog Input/
RTD(S2:3) are not calibrated
Calibrate the Analog Input/RTD input. If
error persists after calibration, contact
Panametrics factory
A7:(S2:4)Ch Not Calibrated
Error occurs when Analog Input/RTD
(S2:4) are not calibrated
Calibrate the Analog Input/RTD input. If
error persists after calibration, contact
Panametrics factory
A10:(S2:3)Input NotConnect!
Analog Input: Error occurs when
(4-20mA) input is not connected at
Channel (S2:3).
RTD Input: Error occurs when RTD input
is not connected or temp greater than
390 deg C at Channel (S2:3)
Check connectivity for Analog Input/
RTD input and RTD temperature.
Try power cycling the meter. If error
persists after power cycle, contact
Panametrics factory
A11:(S2:4)Input NotConnect!
Analog Input: Error occurs when
(4-20mA) input is not connected at
Channel (S2:4).
RTD Input: Error occurs when RTD input
is not connected or temp greater than
390 deg C at Channel (S2:4)
Check connectivity for Analog Input/
RTD input and RTD temperature.
Try power cycling the meter. If error
persists after power cycle, contact
Panametrics factory
A12:(S2:3)Ch OverRange Err!
Exceeds input values. For analog input
(S2:3) greater than 21mA
Ensure analog input current less than
21mA. Try power cycling the meter.
If error persists after power cycle,
contact Panametrics factory
A13:(S2:4)Ch OverRange Err!
Analog input(S2:4) greater than 21mA
Ensure analog input current less than
21mA. Try power cycling the meter.
If error persists after power cycle,
contact Panametrics factory
64
Содержание Panametrics PanaFlow LC
Страница 1: ...PanaFlow LC User s manual 910 327 Rev A...
Страница 2: ...ii...
Страница 3: ...PanaFlow LC Panametrics liquid ultrasonic flowmeter User s manual 910 327 Rev A Sep 2019 iii...
Страница 4: ...no content intended for this page iv...
Страница 12: ...no content intended for this page xii...
Страница 44: ...Figure 58 Setting up pulse output 32...
Страница 49: ...Figure 60 Setting up alarm output 37...
Страница 80: ...no content intended for this page 68...
Страница 82: ...no content intended for this page 70...
Страница 86: ...no content intended for this page 74...
Страница 98: ...Table 30 Service Record Date Description of Service Performed By 86...
Страница 104: ...no content intended for this page 92...
Страница 106: ...no content intended for this page 94...
Страница 114: ...no content intended for this page 102...
Страница 138: ...no content intended for this page 126...
Страница 140: ...no content intended for this page 128...
Страница 145: ...no content intended for this page 133...