Figure 19: Transducer mounting, step 1
2. Apply a thin bead of couplant to one of the transducers.
A bead approximately the size of a toothpaste bead
should be placed down the center of the transducer
face.
Figure 20: Transducer mounting, step 2
Panametrics supplies an ultrasonic couplant for your
clamp-on installation. The purpose of the couplant is
to provide reliable transmission of ultrasound between
two adjacent solid surfaces. Couplants perform this task
by excluding air from the space between the adjacent
surfaces. Accordingly, the clamp-on transducers should be
pressed tightly against the pipe, using hand pressure on
the set screw to squeeze the couplant to as thin a film as
practical for the given pipe surface.
Panametrics provides couplants for both permanent and
temporary use as well as for high-temperature and low-
temperature applications. For long-term installations, make
sure the couplant does not dry out or extrude. Standard
couplants supplied from Panametrics for the
PanaFlow™ LC
are listed in Table 1 below.
Table 1: Standard Panametrics Couplants
Part No.
Type
Temp. Range
Use
CPL-1
Standard
-40 to +240°C Semi-Permanent
CPL-8
Solid Sheet -40 to +260°C
Permanent
3. Determine the upstream and downstream ends of the
pipe and place the appropriate transducer into the
corresponding block on one of the sub-assemblies.
Make sure the transducer cable connector faces away
from the center of the installation.
Figure 21: Transducer mounting on UCF (top left), MCF (top right), GCF (bottom left), and SCF (bottom right)
10
Содержание Panametrics PanaFlow LC
Страница 1: ...PanaFlow LC User s manual 910 327 Rev A...
Страница 2: ...ii...
Страница 3: ...PanaFlow LC Panametrics liquid ultrasonic flowmeter User s manual 910 327 Rev A Sep 2019 iii...
Страница 4: ...no content intended for this page iv...
Страница 12: ...no content intended for this page xii...
Страница 44: ...Figure 58 Setting up pulse output 32...
Страница 49: ...Figure 60 Setting up alarm output 37...
Страница 80: ...no content intended for this page 68...
Страница 82: ...no content intended for this page 70...
Страница 86: ...no content intended for this page 74...
Страница 98: ...Table 30 Service Record Date Description of Service Performed By 86...
Страница 104: ...no content intended for this page 92...
Страница 106: ...no content intended for this page 94...
Страница 114: ...no content intended for this page 102...
Страница 138: ...no content intended for this page 126...
Страница 140: ...no content intended for this page 128...
Страница 145: ...no content intended for this page 133...