completion of the “as found” calibration and identifying the K-Factor values for each calibration point, select [Composite]
table or [Channel] table option.
6. Scroll and highlight [No. of Points] and press [ENTER]. Enter the number of points to enter in the K-Table.
7. Scroll and highlight [K-Table] and press [ENTER]. Select each point and update the [Velocity] or [Reynolds Number] and its
corresponding [K-Factor]. These points define a calibration curve for the XMT1000.
Note:
[K-Table] Velocity or Reynolds number (Point #1 to Point #20) should be entered in ascending order for the meter to
function correctly.
Figure 79: Calibration menu
52
Содержание Panametrics PanaFlow LC
Страница 1: ...PanaFlow LC User s manual 910 327 Rev A...
Страница 2: ...ii...
Страница 3: ...PanaFlow LC Panametrics liquid ultrasonic flowmeter User s manual 910 327 Rev A Sep 2019 iii...
Страница 4: ...no content intended for this page iv...
Страница 12: ...no content intended for this page xii...
Страница 44: ...Figure 58 Setting up pulse output 32...
Страница 49: ...Figure 60 Setting up alarm output 37...
Страница 80: ...no content intended for this page 68...
Страница 82: ...no content intended for this page 70...
Страница 86: ...no content intended for this page 74...
Страница 98: ...Table 30 Service Record Date Description of Service Performed By 86...
Страница 104: ...no content intended for this page 92...
Страница 106: ...no content intended for this page 94...
Страница 114: ...no content intended for this page 102...
Страница 138: ...no content intended for this page 126...
Страница 140: ...no content intended for this page 128...
Страница 145: ...no content intended for this page 133...