Figure 57: Setting up analog output
3.5.2c Calibrating analog output
To calibrate Analog Output, use steps as in section “Log-in and Primary Pages” to navigate to the Input/Output settings page.
You may use a multimeter or DCS/SCADA to calibrate the Analog Output. Regardless of whether multimeter or DCS/SCADA is
used, the steps below remain the same. For better readability, the steps below only indicate multimeter and does not repeat
multimeter or DCS/SCADA.
1. Turn ON the multimeter (if used), and set it to measure Current (mA) DC. Connect test lead from positive side (Aout+) of
the Main Analog Output to the positive terminal of multimeter, and the negative lead to the negative terminal (Aout-).
2. Then in Meter Menu highlight
[Analog Output (std)]
and press
[ENTER]
. Then highlight
[4-20mA]
and press
[ENTER]
.
3. Scroll down and select
[Calibrate]
option.
4. Select
[Calibrate 4mA]
and check if the reading on the multimeter reads 4.00mA ± 0.01mA. If the value on the multimeter
is not 4.00mA ± 0.01mA, input the value read on the multimeter into the Calibrate 4mA value and press
[ENTER]
. Check
multimeter again verify the Current reads 4.00mA within ±0.01mA.
5. Select
[Calibrate 20mA]
and check if the reading on the multimeter reads 20.00mA ± 0.01mA. If the value on the
multimeter is not 20.00mA ± 0.01mA, input the value read on the multimeter into the Calibrate 20mA value and press
[ENTER]
. Check multimeter again verify the Current reads 20.00mA within ±0.01mA.
6. Select
[Percentage of Scale]
and adjust the scale to 0.00% and press
[ENTER]
, then verify the reading on the multimeter
is 4.00mA within ±0.01mA. Then adjust the scale to 50.00% and press
[ENTER]
, then verify the reading on the multimeter is
12.00mA within ±0.01mA. Then adjust the scale to 100.00% and press
[ENTER]
, then verify the reading on the multimeter is
20.00mA within ±0.01mA.
7. If Steps 4, 5, and 6 were successfully completed and verified, the Analog output is successfully calibrated.
30
Содержание Panametrics PanaFlow LC
Страница 1: ...PanaFlow LC User s manual 910 327 Rev A...
Страница 2: ...ii...
Страница 3: ...PanaFlow LC Panametrics liquid ultrasonic flowmeter User s manual 910 327 Rev A Sep 2019 iii...
Страница 4: ...no content intended for this page iv...
Страница 12: ...no content intended for this page xii...
Страница 44: ...Figure 58 Setting up pulse output 32...
Страница 49: ...Figure 60 Setting up alarm output 37...
Страница 80: ...no content intended for this page 68...
Страница 82: ...no content intended for this page 70...
Страница 86: ...no content intended for this page 74...
Страница 98: ...Table 30 Service Record Date Description of Service Performed By 86...
Страница 104: ...no content intended for this page 92...
Страница 106: ...no content intended for this page 94...
Страница 114: ...no content intended for this page 102...
Страница 138: ...no content intended for this page 126...
Страница 140: ...no content intended for this page 128...
Страница 145: ...no content intended for this page 133...