2.11.1 Installing the Junction Box
Mount the aluminum (5lbs) or stainless steel (15lbs) junction
box using the included mounting plate to a wall (hardware
supplied by end user) or a 2 inch pipe (hardware supplied
with J-box). See Figure 40 for reference.
Figure 40: Aluminum J-box mount to 2” pipe using supplied hardware kit
(stainless steel install will be identical)
2.11.2 Single channel set up
Components of a single channel PanaFlow LC will be a
meter head, two transducer cables and a set of transducers
and clamping fixtures. Mount meter head to a two inch post
using included hardware or another location following local
electrical and building codes.
Install transducers as mentioned in Sections 2.7 through
Section 2.9. Prepare to wire transducer cables.
Note:
US/ Canada installations will be required to provide
conduit per their particular location. European installations
will include armored cable.
2.11.2.1 Transducer wiring (ATEX installations)
1. Thread one end of armored cable into the meter head
and the other into the junction box on the back of the
transducer (see Figure 1 on page 1 for reference).
2. Terminate connections of the transducer end of each
cable to each of your clamp on transducers.
3. At the meter end of the cable secure the red wire of
channel one downstream RG62 cable to the CH1 (red
box in Figure 41) DN terminal of the XMT1000 backplane.
4. Secure Black wire of channel one downstream RG62
cable to the CH1 (red box in Figure 41) RTN terminal
directly below the DN terminal of the XMT1000
backplane.
5. Secure red wire of channel one upstream RG62 cable to
the CH1 (red box in Figure 41) UP terminal of the XMT1000
backplane.
6. Secure Black wire of channel one upstream RG62 cable
to the CH1 (red box in Figure 41) RTN terminal directly
above the UP terminal of the XMT1000 backplane.
Figure 41: XMT1000, channel back plane
Figure 42: PanaFlow LC single channel set
2.11.3 Two channel set up
Components of a two channel PanaFlow LC will be a meter
head, 4 transducer cables and two sets of transducers, 2
channel junction box and clamping fixtures. Mount meter
head to a two inch post using included hardware or another
location following local electrical and building codes.
Install transducers as mentioned in Section 2.7 through
Section 2.9. Prepare to wire transducer cables.
Note:
US/ Canada installations will be required to provide
conduit per their particular location. European installations
will include armored cable.
16
Содержание Panametrics PanaFlow LC
Страница 1: ...PanaFlow LC User s manual 910 327 Rev A...
Страница 2: ...ii...
Страница 3: ...PanaFlow LC Panametrics liquid ultrasonic flowmeter User s manual 910 327 Rev A Sep 2019 iii...
Страница 4: ...no content intended for this page iv...
Страница 12: ...no content intended for this page xii...
Страница 44: ...Figure 58 Setting up pulse output 32...
Страница 49: ...Figure 60 Setting up alarm output 37...
Страница 80: ...no content intended for this page 68...
Страница 82: ...no content intended for this page 70...
Страница 86: ...no content intended for this page 74...
Страница 98: ...Table 30 Service Record Date Description of Service Performed By 86...
Страница 104: ...no content intended for this page 92...
Страница 106: ...no content intended for this page 94...
Страница 114: ...no content intended for this page 102...
Страница 138: ...no content intended for this page 126...
Страница 140: ...no content intended for this page 128...
Страница 145: ...no content intended for this page 133...