G.2.4 Emerson wireless gateway setup
configuration
The following information will show how to configure the
Wireless Gateway with a STATIC IP to a dedicated PC, this
can be useful as a diagnostic tool.
Note:
This is not a typical end user application but can be
used as a diagnostic tool at the customer site, the gateway
may already be setup at the customer site per their
procedures.
1. C24VDC power to the wireless Gateway. Refer to
the label on the inside cover of the Gateway.
2. Connect a CROSSOVER cable from the Ethernet 1 port of
the gateway and the Ethernet port on the PC.
3. Configure the PC TCP/IP Settings (Note this information
can be found in the Gateway manual).
a. Open: A network settings in the PC control panel.
b. Select TCP/IP and Properties.
c. Record any current static IP settings, before making
changes so that the PC can be returned to its original
state.
d. Select “Use the following IP Address” box.
e. Input 192.168.1.12 in the IP Address block.
f. Input 255.255.255.0 in the Subnet Mask.
g. Select OK for both TCP/IP and LAN properties Windows.
4. Disable Proxies.
a. Open the standard Web browser on the PC.
b. Navigate to
Tool->Internet Options->Connections->LAN
Settings
.
c. Uncheck the Proxy Server Box.
5. Configure the Wireless Gateway.
a. Under web server, type address https://192.168.1.10.
• Log in as User: admin and Password: default.
b. Navigate to Setup->Internet Protocol Screen.
• Verify the Specify an IP Address box is checked.
• Set IP Address to 192.168.1.10.
• Set mask to 255.255.255.0.
• Set Gateway to 192.168.1.12 (Note: this is the IP Address on
the PC)
6. Verify operation: Note that HART is a slow protocol,
Wireless HART is even slower, so be patient waiting for
the Gateway to see the VECTOR.
a. Look at the Network Overview page to see if there are
any “live” devices on the network.
b. When visible switch to the Explorer page to see the
VECTOR adapter and Wireless HART Data.
104
Содержание Panametrics PanaFlow LC
Страница 1: ...PanaFlow LC User s manual 910 327 Rev A...
Страница 2: ...ii...
Страница 3: ...PanaFlow LC Panametrics liquid ultrasonic flowmeter User s manual 910 327 Rev A Sep 2019 iii...
Страница 4: ...no content intended for this page iv...
Страница 12: ...no content intended for this page xii...
Страница 44: ...Figure 58 Setting up pulse output 32...
Страница 49: ...Figure 60 Setting up alarm output 37...
Страница 80: ...no content intended for this page 68...
Страница 82: ...no content intended for this page 70...
Страница 86: ...no content intended for this page 74...
Страница 98: ...Table 30 Service Record Date Description of Service Performed By 86...
Страница 104: ...no content intended for this page 92...
Страница 106: ...no content intended for this page 94...
Страница 114: ...no content intended for this page 102...
Страница 138: ...no content intended for this page 126...
Страница 140: ...no content intended for this page 128...
Страница 145: ...no content intended for this page 133...