Table 24: System error description and recommended actions
Error Code
Error Message
Description / Recommended Action
S1: In Config Mode
In configuration mode indicator
Indicator:
This is displayed when a user
has logged in to either Operator, Admin
or Factory access level. The indicator
will clear automatically when the user
logs out or saves the configuration
changes
S2: Invalid User
Invalid user warning
Warning:
The passcode entered for
access level is incorrect. Please log
in with the correct access level and
passcode
S3: Invalid Request
Invalid request warning
Warning:
An invalid communication
packet was received and discarded.
Or, the requested operation is invalid.
Please send a valid packet or operation
request
S4: Invalid Param Range
Invalid parameter range warning
Warning:
The value programmed for
the parameter was out of range and
hence discarded. Please enter a valid
range
S5: Unsupported Parameter
This parameter is not supported
Warning:
A read or write request to an
unsupported parameter was received
S6: Flow Measurement
One or more flow measurement
channels are in error
Error:
One or more flow measurement
channels are in error; accuracy of the
measurement may be compromised.
For more details please check flow(E)
errors
S7: Persistent Param CRC
Persistent parameter CRC fault
Fault:
Persistent parameter CRC failed.
Try power cycling the meter. If error
persists after power cycle, contact
Panametrics factory
S11: Clock Frequency
Clock frequency error
Fault:
Input clock frequency failure.
Try power cycling the meter. If error
persists after power cycle, contact
Panametrics factory
S12: CPU
CPU error
Fault:
CPU registers have stuck bits.
Try power cycling the meter. If error
persists after power cycle, contact
Panametrics factory
S13: Invariable Flash Memory
Flash memory fault
Fault:
Flash memory test failed. Try
power cycling the meter. If error
persists after power cycle, contact
Panametrics factory
60
Содержание Panametrics PanaFlow LC
Страница 1: ...PanaFlow LC User s manual 910 327 Rev A...
Страница 2: ...ii...
Страница 3: ...PanaFlow LC Panametrics liquid ultrasonic flowmeter User s manual 910 327 Rev A Sep 2019 iii...
Страница 4: ...no content intended for this page iv...
Страница 12: ...no content intended for this page xii...
Страница 44: ...Figure 58 Setting up pulse output 32...
Страница 49: ...Figure 60 Setting up alarm output 37...
Страница 80: ...no content intended for this page 68...
Страница 82: ...no content intended for this page 70...
Страница 86: ...no content intended for this page 74...
Страница 98: ...Table 30 Service Record Date Description of Service Performed By 86...
Страница 104: ...no content intended for this page 92...
Страница 106: ...no content intended for this page 94...
Страница 114: ...no content intended for this page 102...
Страница 138: ...no content intended for this page 126...
Страница 140: ...no content intended for this page 128...
Страница 145: ...no content intended for this page 133...