Figure 69: Fluid programming
3.6.3 Programming the Path Configuration
Use steps as in section “Log-in and Primary Pages” to navigate to the Programming page. Refer to Figure 70 for the Path
configuration options.
1. Highlight [Composite] and press [ENTER]. Scroll down and select [Path] and press [ENTER].
2. Select [Path Configuration], [Path Weights] and [Path Error Handling].
3. Path Weights are used in Composite flow velocity calculations as in the following equation:
4. If [Path Error Handling] is set to On, the meter will continue to provide measurements even if one or two channels are in
error. Unless, all three channels (for 3 Path (TD-TD-TD)) are in error the flow measurement continues.
Figure 70: Path configuration
46
Содержание Panametrics PanaFlow LC
Страница 1: ...PanaFlow LC User s manual 910 327 Rev A...
Страница 2: ...ii...
Страница 3: ...PanaFlow LC Panametrics liquid ultrasonic flowmeter User s manual 910 327 Rev A Sep 2019 iii...
Страница 4: ...no content intended for this page iv...
Страница 12: ...no content intended for this page xii...
Страница 44: ...Figure 58 Setting up pulse output 32...
Страница 49: ...Figure 60 Setting up alarm output 37...
Страница 80: ...no content intended for this page 68...
Страница 82: ...no content intended for this page 70...
Страница 86: ...no content intended for this page 74...
Страница 98: ...Table 30 Service Record Date Description of Service Performed By 86...
Страница 104: ...no content intended for this page 92...
Страница 106: ...no content intended for this page 94...
Страница 114: ...no content intended for this page 102...
Страница 138: ...no content intended for this page 126...
Страница 140: ...no content intended for this page 128...
Страница 145: ...no content intended for this page 133...