Calibration and Troubleshooting
M370430-01 Rev B
7-25
7
Storing and Loading Calibration Parameters
It is recommended that you save all the calibration parameters in a
text file so they can be reloaded in the event of a flash failure or a
mistake in calibration.
To get the calibration data, execute the following commands and
save the response in a text file:
The SCPI command (s) to download all the calibration data are:
[:]CALibration:PARameter[:OUTPut]
[:]CALibration:PARameter:APRogram?
[:]CALibration:PARameter:INPut?
The response to these commands are in the 488.2 Arbitrary
Program Block Data format.
If the calibration ever needs to be restored, you can execute the
same commands and write the arbitrary block data response that
was received in the query as the parameter.
For example, downloading the parameter data can be done from
the hyper terminal as follows.
:cal:par?
#2 62#h2710,#h0000,#h2710,#h0000,#h2710,#h0000,#h2710,
#h0000,#h1AA4
:cal:par:apr?
#259#h7B,#h7F,#h7F,#h7F,#h7F,#h7F,#h7F,#h7F,#h7F,#h7F,
#h 7F, #h7F
:cal:par:inp?
#227#h2710,#h00,#h2710,#h00
To restore the unit with a set of previously saved calibration values
can be done as follows.
:cal:par
#262#h2710,#h0000,#h2710,#h0000,#h2710,#h0000,#h2710,#h0
000,#h1AA4
:cal:par:apr
#259#h7F,#h7F,#h7F,#h7F,#h7F,#h7F,#h7F,#h7F,#h7F,#h7F,#h7
F,#h7F
:cal:par:inp
#h227#h2710,#h00,#h2710,#h00
By saving and reloading the calibration parameters using these
commands, you can calibrate the unit for maximum accuracy for a
specific load and then switch loads and load a calibration set that is
accurate at maximum accuracy for the new load.
Содержание Sorensen XG
Страница 2: ......
Страница 22: ...Contents xx M370430 01 Rev B ...
Страница 23: ...1 Introduction Chapter 1 Introduction describes the features of the XG XTR Family Programmable DC Power Supply ...
Страница 94: ...Local Operation 3 46 M370430 01 Rev B Figure 3 4 Split Supply Operation Power Supply Power Supply Common Load ...
Страница 96: ...Local Operation 3 48 M370430 01 Rev B Figure 3 5 Series Operation ...
Страница 102: ...Local Operation 3 54 M370430 01 Rev B ...
Страница 210: ...Remote Operation 5 76 M370430 01 Rev B ...
Страница 248: ...Communications Options 6 38 M370430 01 Rev B 9 In the Setup Type window choose Typical and click Next ...
Страница 249: ...Communications Options M370430 01 Rev B 6 39 6 10 Check that your Settings are correct and then click Install ...
Страница 256: ...Communications Options 6 46 M370430 01 Rev B 9 In the Setup Type window choose Typical and click Next ...
Страница 276: ...Communications Options 6 66 M370430 01 Rev B ...
Страница 358: ...Specifications C 12 M370430 01 Rev B Figure C 2 XG XTR 850 Watt Mechanical Dimensions 60 V to 600 V Models ...
Страница 359: ...Specifications M370430 01 Rev B C 13 Figure C 3 XG XTR in Full Rack Mechanical Dimensions 6 V to 40 V Models ...
Страница 361: ...Specifications M370430 01 Rev B C 15 Figure C 5 XG XTR in Full Rack Mechanical Dimensions 60 V to 600 V Models ...
Страница 364: ...Specifications C 18 M370430 01 Rev B Figure C 9 XG850 Low Voltage Model Dimensions Top Side and Rear ...
Страница 365: ...Specifications M370430 01 Rev B C 19 Figure C 10 XG850 Medium and High Voltage Models Dimensions Top and Rear ...
Страница 371: ...Rack Mount Options and Installation Instructions M370430 01 Rev B D 5 Figure D 3 RM XG2 RM XTR2 Rack Rails ...
Страница 377: ...Rack Mount Options and Installation Instructions M370430 01 Rev B D 11 Figure D 10 Installing Chassis ...
Страница 378: ...Rack Mount Options and Installation Instructions D 12 M370430 01 Rev B ...
Страница 381: ......