Calibration and Troubleshooting
M370430-01 Rev B
7-21
7
Isolated Resistive Programming of Current Calibration
To calibrate the isolated resistive programming of current:
1.
Short the main output with a shunt. Connect a multimeter to
measure the voltage across the shunt.
2.
Connect the 4.000 kΩ resistor to the isolated connector. One
terminal to resistive programming of current (J3.7), and
voltage programming of current (J3.3), and the other to
common (J3.2).
3.
Set the Resistive APG level to 4 kΩ The SCPI command is:
SYST:COMM:APR:LEV:CURR 4
4.
Set the input source as Resistive Isolated for Current
Programming. The SCPI command is:
SYST:REM:SOUR:CURR IAR
5.
Turn the main output On or send the command:
:OUTP ON
6.
Enter the interactive calibration mode by sending the SCPI
command:
CAL:OUTP:ANAL:CURR:RES:ISOL
7.
Press the terminal keys - or + to reduce or increase the main
output current value. Adjust the main output current to 100%.
8.
Press any key other than the + or - keys to exit from the
interactive calibration mode.
Содержание Sorensen XG
Страница 2: ......
Страница 22: ...Contents xx M370430 01 Rev B ...
Страница 23: ...1 Introduction Chapter 1 Introduction describes the features of the XG XTR Family Programmable DC Power Supply ...
Страница 94: ...Local Operation 3 46 M370430 01 Rev B Figure 3 4 Split Supply Operation Power Supply Power Supply Common Load ...
Страница 96: ...Local Operation 3 48 M370430 01 Rev B Figure 3 5 Series Operation ...
Страница 102: ...Local Operation 3 54 M370430 01 Rev B ...
Страница 210: ...Remote Operation 5 76 M370430 01 Rev B ...
Страница 248: ...Communications Options 6 38 M370430 01 Rev B 9 In the Setup Type window choose Typical and click Next ...
Страница 249: ...Communications Options M370430 01 Rev B 6 39 6 10 Check that your Settings are correct and then click Install ...
Страница 256: ...Communications Options 6 46 M370430 01 Rev B 9 In the Setup Type window choose Typical and click Next ...
Страница 276: ...Communications Options 6 66 M370430 01 Rev B ...
Страница 358: ...Specifications C 12 M370430 01 Rev B Figure C 2 XG XTR 850 Watt Mechanical Dimensions 60 V to 600 V Models ...
Страница 359: ...Specifications M370430 01 Rev B C 13 Figure C 3 XG XTR in Full Rack Mechanical Dimensions 6 V to 40 V Models ...
Страница 361: ...Specifications M370430 01 Rev B C 15 Figure C 5 XG XTR in Full Rack Mechanical Dimensions 60 V to 600 V Models ...
Страница 364: ...Specifications C 18 M370430 01 Rev B Figure C 9 XG850 Low Voltage Model Dimensions Top Side and Rear ...
Страница 365: ...Specifications M370430 01 Rev B C 19 Figure C 10 XG850 Medium and High Voltage Models Dimensions Top and Rear ...
Страница 371: ...Rack Mount Options and Installation Instructions M370430 01 Rev B D 5 Figure D 3 RM XG2 RM XTR2 Rack Rails ...
Страница 377: ...Rack Mount Options and Installation Instructions M370430 01 Rev B D 11 Figure D 10 Installing Chassis ...
Страница 378: ...Rack Mount Options and Installation Instructions D 12 M370430 01 Rev B ...
Страница 381: ......