Local Operation
M370430-01 Rev B
3-31
3
Defining the UVP Set Point
To define the UVP set point:
1.
Turn the power supply On.
2.
Set the output to the desired voltage.
3.
Turn the 9-position mode control to the PRT position.
is displayed.
4.
Turn the rotary Adjust/Enter control until
is
displayed.
5.
Press the rotary Adjust/Enter control to select the UVP for
setting up.
6.
Set up the lower limit protection using the rotary Adjust/Enter
control.
Important:
The UVP range is variable from 0 volts to OVP
SET
-10%.
7.
Once the desired value has been set, press the rotary
Adjust/Enter control to commit the setting.
See “Coarse and Fine Adjustment Modes” on page 3-3.
The SCPI command (s) for these instructions are:
[[:]SOURce[<channel>]]:VOLTage:PROTection:UNDer[:LEVel]
Over Current Protection (OCP)
The OCP alarm is a firmware alarm that is only configurable using the
SCPI interface. The OCP triggers if the output current exceeds the OCP
set point. Because the OCP alarm is a firmware based alarm it has a
latency before detection of the condition can occur. The user should be
aware of this latency and take this into consideration when relying on
OCP to protect a sensitive load. The maximum period for detecting an
OCP condition is 300 ms.
OCP is fixed at 110% of rated current.
Содержание Sorensen XG
Страница 2: ......
Страница 22: ...Contents xx M370430 01 Rev B ...
Страница 23: ...1 Introduction Chapter 1 Introduction describes the features of the XG XTR Family Programmable DC Power Supply ...
Страница 94: ...Local Operation 3 46 M370430 01 Rev B Figure 3 4 Split Supply Operation Power Supply Power Supply Common Load ...
Страница 96: ...Local Operation 3 48 M370430 01 Rev B Figure 3 5 Series Operation ...
Страница 102: ...Local Operation 3 54 M370430 01 Rev B ...
Страница 210: ...Remote Operation 5 76 M370430 01 Rev B ...
Страница 248: ...Communications Options 6 38 M370430 01 Rev B 9 In the Setup Type window choose Typical and click Next ...
Страница 249: ...Communications Options M370430 01 Rev B 6 39 6 10 Check that your Settings are correct and then click Install ...
Страница 256: ...Communications Options 6 46 M370430 01 Rev B 9 In the Setup Type window choose Typical and click Next ...
Страница 276: ...Communications Options 6 66 M370430 01 Rev B ...
Страница 358: ...Specifications C 12 M370430 01 Rev B Figure C 2 XG XTR 850 Watt Mechanical Dimensions 60 V to 600 V Models ...
Страница 359: ...Specifications M370430 01 Rev B C 13 Figure C 3 XG XTR in Full Rack Mechanical Dimensions 6 V to 40 V Models ...
Страница 361: ...Specifications M370430 01 Rev B C 15 Figure C 5 XG XTR in Full Rack Mechanical Dimensions 60 V to 600 V Models ...
Страница 364: ...Specifications C 18 M370430 01 Rev B Figure C 9 XG850 Low Voltage Model Dimensions Top Side and Rear ...
Страница 365: ...Specifications M370430 01 Rev B C 19 Figure C 10 XG850 Medium and High Voltage Models Dimensions Top and Rear ...
Страница 371: ...Rack Mount Options and Installation Instructions M370430 01 Rev B D 5 Figure D 3 RM XG2 RM XTR2 Rack Rails ...
Страница 377: ...Rack Mount Options and Installation Instructions M370430 01 Rev B D 11 Figure D 10 Installing Chassis ...
Страница 378: ...Rack Mount Options and Installation Instructions D 12 M370430 01 Rev B ...
Страница 381: ......