Local Operation
M370430-01 Rev B
3-15
3
The reverse operating mode change can also occur if the load resistance
is increased to the point that the required load current drops below the
I
SET
value. At that point the power supply would crossover to CV mode
and the load current would be free to vary as the load resistance
changed.
L
o
a
d
c
u
rr
e
n
t (
V
)
L
o
a
d
c
u
rr
e
n
t (
A
)
Iset
Figure 3-3
Operating Modes
To set the voltage set point (VSET)
:
1.
Turn the 9-position mode control to the VOLTS position or
press the rotary Adjust/Enter control if the 9-position mode
control is already at the VOLTS position.
The voltage set point is blinking in the output voltage display. The
output current will be displayed in the output current display.
2.
Set the desired voltage value using the rotary Adjust/Enter
control.
Important:
If you can't get the set point to the desired level, you may need to
change the set points for OVP.
3.
Press the rotary Adjust/Enter control to transition to Fine
Adjust mode.
The set point will blink faster to indicate fine adjust tracking mode.
4.
Set the desired voltage value using the rotary Adjust/Enter
control.
CV mode
Crossover point
CV mode
Vset
Load current
Load voltage
Load resistance (
Ω
Содержание Sorensen XG
Страница 2: ......
Страница 22: ...Contents xx M370430 01 Rev B ...
Страница 23: ...1 Introduction Chapter 1 Introduction describes the features of the XG XTR Family Programmable DC Power Supply ...
Страница 94: ...Local Operation 3 46 M370430 01 Rev B Figure 3 4 Split Supply Operation Power Supply Power Supply Common Load ...
Страница 96: ...Local Operation 3 48 M370430 01 Rev B Figure 3 5 Series Operation ...
Страница 102: ...Local Operation 3 54 M370430 01 Rev B ...
Страница 210: ...Remote Operation 5 76 M370430 01 Rev B ...
Страница 248: ...Communications Options 6 38 M370430 01 Rev B 9 In the Setup Type window choose Typical and click Next ...
Страница 249: ...Communications Options M370430 01 Rev B 6 39 6 10 Check that your Settings are correct and then click Install ...
Страница 256: ...Communications Options 6 46 M370430 01 Rev B 9 In the Setup Type window choose Typical and click Next ...
Страница 276: ...Communications Options 6 66 M370430 01 Rev B ...
Страница 358: ...Specifications C 12 M370430 01 Rev B Figure C 2 XG XTR 850 Watt Mechanical Dimensions 60 V to 600 V Models ...
Страница 359: ...Specifications M370430 01 Rev B C 13 Figure C 3 XG XTR in Full Rack Mechanical Dimensions 6 V to 40 V Models ...
Страница 361: ...Specifications M370430 01 Rev B C 15 Figure C 5 XG XTR in Full Rack Mechanical Dimensions 60 V to 600 V Models ...
Страница 364: ...Specifications C 18 M370430 01 Rev B Figure C 9 XG850 Low Voltage Model Dimensions Top Side and Rear ...
Страница 365: ...Specifications M370430 01 Rev B C 19 Figure C 10 XG850 Medium and High Voltage Models Dimensions Top and Rear ...
Страница 371: ...Rack Mount Options and Installation Instructions M370430 01 Rev B D 5 Figure D 3 RM XG2 RM XTR2 Rack Rails ...
Страница 377: ...Rack Mount Options and Installation Instructions M370430 01 Rev B D 11 Figure D 10 Installing Chassis ...
Страница 378: ...Rack Mount Options and Installation Instructions D 12 M370430 01 Rev B ...
Страница 381: ......