5.2. Specifications
Infant
Pediatric
Adult
Resuscitator volume
approx. 220 ml
approx. 670 ml
approx. 1510 ml
Delivered volume one hand*
150 ml
450 ml
600 ml
Delivered volume two hands*
-
-
1000 ml
Dimensions
(length x diameter) w/o
reservoir and accessory
approx.
190 x 71 mm
approx.
223 x 99 mm
approx.
284 x 127 mm
Weight w/o reservoir
and accessory
approx. 70 g
approx. 145 g
approx. 220 g
Pressure-limiting valve**
4.0 kPa
(40 cm H
2
O)
4.0 kPa
(40 cm H
2
O)
4.0 kPa
(40 cm H
2
O)
Dead space
≤ 5 ml + 10 % of the
delivered volume
≤ 5 ml + 10 % of the
delivered volume
≤ 5 ml + 10 % of the
delivered volume
Inspiratory resistance***
max 0.1 kPa
(1.0 cm H
2
O)
at 5 l/min
max 0.5 kPa
(5.0 cm H
2
O)
at 50 l/min
max 0.5 kPa
(5.0 cm H
2
O)
at 50 l/min
Expiratory resistance***
max 0.2 kPa
(2.0 cm H
2
O)
at 5 l/min
max 0.27 kPa
(2.7 cm H
2
O)
at 50 l/min
max 0.27 kPa
(2.7 cm H
2
O)
at 50 l/min
Reservoir volume
approx. 300 ml (bag)
approx. 100 ml (tube)
approx. 2600 ml (bag) approx. 2600 ml (bag)
Patient connector
Outside 22 mm male (ISO 5356-1)
Inside 15 mm female (ISO 5356-1)
Expiration connector
(for PEEP valve attachment)
30 mm male (ISO 5356-1)
Manometer Port connector
Ø 4.2 +/- 0.1 mm
Demand Valve connector
-
Inside 32 mm female (EN ISO 10651-4)
Infant
Pediatric
Adult
Forward and
backward leakage
Not measurable
M-Port
Connector compatible with EN ISO 80369-7
O
2
inlet connector
According to EN 13544-2
Operation temperature limits -18 °C to +50 °C (-0.4 °F to +122 °F), tested according to EN ISO 10651-4
Storage temperature limits
-40 °C to +60 °C (-40 °F to +140 °F), tested according to EN ISO 10651-4
Recommended long term storage in closed packaging at room temperature, away from sunlight.
* Tested according to EN ISO 10651-4.
** Higher delivery pressure can be obtained by overriding the pressure-limiting valve.
*** At general test conditions according to EN ISO 10651-4.
5.3. MRI Safety Information
The Ambu SPUR II Resuscitator
1
, and Ambu SPUR II Resuscitator with attached Ambu
Disposable PEEP 20 Valve
2
, and Ambu SPUR II Resuscitator with attached Ambu Disposable
PEEP 20 Valve and Ambu Disposable Pressure Manometer
2
are MR Conditional, and
therefore may be safely used in the MR environment (not inside the MR bore) under the
following conditions.
Static magnetic field of 7 Tesla and less, with
• Maximum spatial field gradient of
– 10,000 G/cm (100 T/m)
1
– 16,000 G/cm (160 T/m)
2
• Maximum force product of
– 450,000,000 G2/cm (450 T
2
/m)
1
– 721,000,000 G2/cm (721 T
2
/m)
2
Use inside the MR bore may influence MR image quality.
RF-induced heating and MR image artifacts have not been tested. Any metallic parts are
fully encapsulated and do not have any contact with the human body.
Содержание SPUR II
Страница 1: ...INSTRUCTIONS FOR USE Ambu SPUR II Disposable...
Страница 16: ...16 4 2 2 2 2 1 2 2 0 C 0 C 10 l min 10 l min M Port M Port M Port 2 3 4 3 3 1 3 2...
Страница 46: ...46 25 0 C 0 C 26 1 2 3 4 EtCO2 5 6 Ambu SPUR II Ambu SPUR II 7 1 8 1 9 2 Ambu SPUR II Ambu PEEP Ambu Ambu 4 3...
Страница 48: ...48 1 5 PEEP 4 2 4 2 1 4 2 2 0 C 4 2 2 2 2 1 2 2 0 C 0 C 10 l min 10 l min 2 3 4 3 3 1...
Страница 108: ...108 1 2 3 4 M EtCO2 5 6 7 1 8 1 9 2 Ambu SPUR II 4 3 Ambu Ambu PEEP Ambu 3 30 kg 6 30 kg 10 kg MRI...
Страница 110: ...110 2 2 0 C 0 C 10 l 2 10 l M M M 2 3 4 3 3 1 3 2 4 40 cm H2 O 4 0kPa 4 1 40 cm H2 O 4 2 M Ambu SPUR II M M M M EtCO2 M...
Страница 170: ...170 1 4 1 2 3 Ambu PEEP 4 5 6 7 8 9 10 11 CO2 12 O2 13 EtCO2 400 EtCO2 14 1 15 O2 16 17 18 19...
Страница 171: ...ru 171 20 21 22 PEEP 23 Ambu SPUR II 24 O2 25 0 C 0 C 26 1 2 3 4 M EtCO2 5 6 Ambu SPUR II Ambu SPUR II...
Страница 172: ...172 7 1 8 1 9 2 Ambu SPUR II Ambu Ambu PEEP Ambu 4 3 3 30 6 30 10 CN https www ambu com symbol explanation...
Страница 173: ...ru 173 4 4 1 1 1a 1b 1c 1d EtCO2 1e 1 1 1 2 1 3 1 4 1 5 1 6 1 7 1 8 1 5 PEEP 4 2 4 2 1 4 2 2 0 C 4 2 2 2 2 1 2 2...
Страница 174: ...174 0 C 0 C 10 10 2 3 4 3 3 1 3 2 Ambu Disposable Pressure Manometer 4 40 H2 O 4 0 4 1 40 H2 O 4 2...
Страница 216: ...492 2300 40 V11 2022 04 Ambu A S TCC 11130 Ambu A S Baltorpbakken 13 2750 Ballerup Denmark T 45 72 25 20 00 ambu com...