Acquisition Control
12
Agilent InfiniiVision 2000 X-Series Oscilloscopes User's Guide
155
Overview of Sampling
To understand the oscilloscope's sampling and acquisition modes, it is
helpful to understand sampling theory, aliasing, oscilloscope bandwidth
and sample rate, oscilloscope rise time, oscilloscope bandwidth required,
and how memory depth affects sample rate.
Sampling Theory
The Nyquist sampling theorem states that for a limited bandwidth
(band- limited) signal with maximum frequency f
MAX
, the equally spaced
sampling frequency f
S
must be greater than twice the maximum frequency
f
MAX
, in order to have the signal be uniquely reconstructed without
aliasing.
f
MAX
= f
S
/2 = Nyquist frequency (f
N
) = folding frequency
Aliasing
Aliasing occurs when signals are under- sampled (f
S
< 2f
MAX
). Aliasing is
the signal distortion caused by low frequencies falsely reconstructed from
an insufficient number of sample points.
Figure 21
Aliasing
Содержание 2000 X-Series
Страница 1: ...s1 Agilent InfiniiVision 2000 X Series Oscilloscopes User s Guide ...
Страница 20: ...20 Agilent InfiniiVision 2000 X Series Oscilloscopes User s Guide ...
Страница 44: ...44 Agilent InfiniiVision 2000 X Series Oscilloscopes User s Guide 1 Getting Started ...
Страница 58: ...58 Agilent InfiniiVision 2000 X Series Oscilloscopes User s Guide 2 Horizontal Controls ...
Страница 66: ...66 Agilent InfiniiVision 2000 X Series Oscilloscopes User s Guide 3 Vertical Controls ...
Страница 80: ...80 Agilent InfiniiVision 2000 X Series Oscilloscopes User s Guide 4 Math Waveforms ...
Страница 104: ...104 Agilent InfiniiVision 2000 X Series Oscilloscopes User s Guide 6 Digital Channels ...
Страница 110: ...110 Agilent InfiniiVision 2000 X Series Oscilloscopes User s Guide 7 Serial Decode ...
Страница 116: ...116 Agilent InfiniiVision 2000 X Series Oscilloscopes User s Guide 8 Display Settings ...
Страница 152: ...152 Agilent InfiniiVision 2000 X Series Oscilloscopes User s Guide 11 Trigger Mode Coupling ...
Страница 178: ...178 Agilent InfiniiVision 2000 X Series Oscilloscopes User s Guide 13 Cursors ...
Страница 214: ...214 Agilent InfiniiVision 2000 X Series Oscilloscopes User s Guide 16 Digital Voltmeter ...
Страница 226: ...226 Agilent InfiniiVision 2000 X Series Oscilloscopes User s Guide 17 Waveform Generator ...
Страница 238: ...238 Agilent InfiniiVision 2000 X Series Oscilloscopes User s Guide 18 Save Recall Setups Screens Data ...
Страница 244: ...244 Agilent InfiniiVision 2000 X Series Oscilloscopes User s Guide 19 Print Screens ...
Страница 280: ...280 Agilent InfiniiVision 2000 X Series Oscilloscopes User s Guide 21 Web Interface ...
Страница 334: ...334 Agilent InfiniiVision 2000 X Series Oscilloscopes User s Guide 24 I2C SPI Triggering and Serial Decode ...
Страница 352: ...352 Agilent InfiniiVision 2000 X Series Oscilloscopes User s Guide Index ...