Horizontal Controls
2
Agilent InfiniiVision 2000 X-Series Oscilloscopes User's Guide
47
Notice how the time/div information in the status line changes.
The
∇
symbol at the top of the display indicates the time reference point.
The horizontal scale knob works (in the Normal time mode) while
acquisitions are running or when they are stopped. When running,
adjusting the horizontal scale knob changes the sample rate. When
stopped, adjusting the horizontal scale knob lets you zoom into acquired
data. See
"Panning and Zooming Single or Stopped Acquisitions"
on
page 48.
Note that the horizontal scale knob has a different purpose in the Zoom
display. See
"To display the zoomed time base"
on page 52.
To adjust the horizontal delay (position)
1
Turn the horizontal delay (position) knob (
).
The trigger point moves horizontally, pausing at 0.00 s (mimicking a
mechanical detent), and the delay value is displayed in the status line.
Changing the delay time moves the trigger point (solid inverted triangle)
horizontally and indicates how far it is from the time reference point
(hollow inverted triangle
∇
). These reference points are indicated along
the top of the display grid.
Figure 2
shows the trigger point with the delay time set to 200 µs. The
delay time number tells you how far the time reference point is located
from the trigger point. When delay time is set to zero, the delay time
indicator overlays the time reference indicator.
All events displayed left of the trigger point happened before the trigger
occurred. These events are called pre- trigger information, and they show
events that led up to the trigger point.
Everything to the right of the trigger point is called post- trigger
information. The amount of delay range (pre- trigger and post- trigger
information) available depends on the time/div selected and memory
depth.
The horizontal position knob works (in the Normal time mode) while
acquisitions are running or when they are stopped. When running,
adjusting the horizontal scale knob changes the sample rate. When
Содержание 2000 X-Series
Страница 1: ...s1 Agilent InfiniiVision 2000 X Series Oscilloscopes User s Guide ...
Страница 20: ...20 Agilent InfiniiVision 2000 X Series Oscilloscopes User s Guide ...
Страница 44: ...44 Agilent InfiniiVision 2000 X Series Oscilloscopes User s Guide 1 Getting Started ...
Страница 58: ...58 Agilent InfiniiVision 2000 X Series Oscilloscopes User s Guide 2 Horizontal Controls ...
Страница 66: ...66 Agilent InfiniiVision 2000 X Series Oscilloscopes User s Guide 3 Vertical Controls ...
Страница 80: ...80 Agilent InfiniiVision 2000 X Series Oscilloscopes User s Guide 4 Math Waveforms ...
Страница 104: ...104 Agilent InfiniiVision 2000 X Series Oscilloscopes User s Guide 6 Digital Channels ...
Страница 110: ...110 Agilent InfiniiVision 2000 X Series Oscilloscopes User s Guide 7 Serial Decode ...
Страница 116: ...116 Agilent InfiniiVision 2000 X Series Oscilloscopes User s Guide 8 Display Settings ...
Страница 152: ...152 Agilent InfiniiVision 2000 X Series Oscilloscopes User s Guide 11 Trigger Mode Coupling ...
Страница 178: ...178 Agilent InfiniiVision 2000 X Series Oscilloscopes User s Guide 13 Cursors ...
Страница 214: ...214 Agilent InfiniiVision 2000 X Series Oscilloscopes User s Guide 16 Digital Voltmeter ...
Страница 226: ...226 Agilent InfiniiVision 2000 X Series Oscilloscopes User s Guide 17 Waveform Generator ...
Страница 238: ...238 Agilent InfiniiVision 2000 X Series Oscilloscopes User s Guide 18 Save Recall Setups Screens Data ...
Страница 244: ...244 Agilent InfiniiVision 2000 X Series Oscilloscopes User s Guide 19 Print Screens ...
Страница 280: ...280 Agilent InfiniiVision 2000 X Series Oscilloscopes User s Guide 21 Web Interface ...
Страница 334: ...334 Agilent InfiniiVision 2000 X Series Oscilloscopes User s Guide 24 I2C SPI Triggering and Serial Decode ...
Страница 352: ...352 Agilent InfiniiVision 2000 X Series Oscilloscopes User s Guide Index ...