114
Agilent InfiniiVision 2000 X-Series Oscilloscopes User's Guide
8
Display Settings
3
To erase the results of previous acquisitions from the display, press the
Clear Persistence
softkey.
The oscilloscope will start to accumulate acquisitions again.
4
To return the oscilloscope to the normal display mode, turn off
persistence; then, press the
Clear Persistence
softkey.
Turning off persistence does not clear the display. The display is
cleared if you press the
Clear Display
softkey or if you press the
[AutoScale]
key (which also turns off persistence).
For another method of seeing worst- case extremes of varying waveforms,
see
"Glitch or Narrow Pulse Capture"
on page 161.
To clear the display
1
Press
[Display] > Clear Display
.
You can also configure the
[Quick Action]
key to clear the display. See
"Configuring the [Quick Action] Key"
on page 261.
To select the grid type
When the
Video
trigger type is selected (see
"Video Trigger"
on page 133),
and the vertical scaling of at least one displayed channel is 140 mV/div,
the
Grid
softkey lets you select from these grid types:
•
Full
— the normal oscilloscope grid.
•
mV
— shows vertical grids, labeled on the left, from - 0.3 V to 0.8 V.
•
IRE
— (Institute of Radio Engineers) shows vertical grids in IRE units,
labeled on the left, from - 40 to 100 IRE. The 0.35 V and 0.7 V levels
from the
mV
grid are also shown and labeled at the right. When the
IRE
grid is selected, cursor values are also shown in IRE units. (Cursor
values via the remote interface are not in IRE units.)
The
mV
and
IRE
grid values are accurate (and match Y cursor values)
when the vertical scaling is 140 mV/division and the vertical offset is
245 mV.
To select the grid type:
Содержание 2000 X-Series
Страница 1: ...s1 Agilent InfiniiVision 2000 X Series Oscilloscopes User s Guide ...
Страница 20: ...20 Agilent InfiniiVision 2000 X Series Oscilloscopes User s Guide ...
Страница 44: ...44 Agilent InfiniiVision 2000 X Series Oscilloscopes User s Guide 1 Getting Started ...
Страница 58: ...58 Agilent InfiniiVision 2000 X Series Oscilloscopes User s Guide 2 Horizontal Controls ...
Страница 66: ...66 Agilent InfiniiVision 2000 X Series Oscilloscopes User s Guide 3 Vertical Controls ...
Страница 80: ...80 Agilent InfiniiVision 2000 X Series Oscilloscopes User s Guide 4 Math Waveforms ...
Страница 104: ...104 Agilent InfiniiVision 2000 X Series Oscilloscopes User s Guide 6 Digital Channels ...
Страница 110: ...110 Agilent InfiniiVision 2000 X Series Oscilloscopes User s Guide 7 Serial Decode ...
Страница 116: ...116 Agilent InfiniiVision 2000 X Series Oscilloscopes User s Guide 8 Display Settings ...
Страница 152: ...152 Agilent InfiniiVision 2000 X Series Oscilloscopes User s Guide 11 Trigger Mode Coupling ...
Страница 178: ...178 Agilent InfiniiVision 2000 X Series Oscilloscopes User s Guide 13 Cursors ...
Страница 214: ...214 Agilent InfiniiVision 2000 X Series Oscilloscopes User s Guide 16 Digital Voltmeter ...
Страница 226: ...226 Agilent InfiniiVision 2000 X Series Oscilloscopes User s Guide 17 Waveform Generator ...
Страница 238: ...238 Agilent InfiniiVision 2000 X Series Oscilloscopes User s Guide 18 Save Recall Setups Screens Data ...
Страница 244: ...244 Agilent InfiniiVision 2000 X Series Oscilloscopes User s Guide 19 Print Screens ...
Страница 280: ...280 Agilent InfiniiVision 2000 X Series Oscilloscopes User s Guide 21 Web Interface ...
Страница 334: ...334 Agilent InfiniiVision 2000 X Series Oscilloscopes User s Guide 24 I2C SPI Triggering and Serial Decode ...
Страница 352: ...352 Agilent InfiniiVision 2000 X Series Oscilloscopes User s Guide Index ...