132
Agilent InfiniiVision 2000 X-Series Oscilloscopes User's Guide
10
Triggers
•
For digital channels, press the
[Digital]
key and select
Thresholds
to set
the threshold level.
The value of the trigger level or digital threshold is displayed in the
upper- right corner of the display.
5
Press the pulse polarity softkey to select positive (
) or negative (
)
polarity for the pulse width you want to capture.
The selected pulse polarity is displayed in the upper- right corner of the
display. A positive pulse is higher than the current trigger level or
threshold and a negative pulse is lower than the current trigger level or
threshold.
When triggering on a positive pulse, the trigger will occur on the high
to low transition of the pulse if the qualifying condition is true. When
triggering on a negative pulse, the trigger will occur on the low to high
transition of the pulse if the qualifying condition is true.
6
Press the qualifier softkey (
< > ><
) to select the time qualifier.
The Qualifier softkey can set the oscilloscope to trigger on a pulse
width that is:
•
Less than a time value (
<
).
For example, for a positive pulse, if you set t<10 ns:
•
Greater than a time value (
>
).
For example, for a positive pulse, if you set t>10 ns:
•
Within a range of time values (
><
).
For example, for a positive pulse, if you set
t>10 ns and t<15 ns:
10 ns
10 ns
Trigger
10 ns
10 ns
Trigger
Содержание 2000 X-Series
Страница 1: ...s1 Agilent InfiniiVision 2000 X Series Oscilloscopes User s Guide ...
Страница 20: ...20 Agilent InfiniiVision 2000 X Series Oscilloscopes User s Guide ...
Страница 44: ...44 Agilent InfiniiVision 2000 X Series Oscilloscopes User s Guide 1 Getting Started ...
Страница 58: ...58 Agilent InfiniiVision 2000 X Series Oscilloscopes User s Guide 2 Horizontal Controls ...
Страница 66: ...66 Agilent InfiniiVision 2000 X Series Oscilloscopes User s Guide 3 Vertical Controls ...
Страница 80: ...80 Agilent InfiniiVision 2000 X Series Oscilloscopes User s Guide 4 Math Waveforms ...
Страница 104: ...104 Agilent InfiniiVision 2000 X Series Oscilloscopes User s Guide 6 Digital Channels ...
Страница 110: ...110 Agilent InfiniiVision 2000 X Series Oscilloscopes User s Guide 7 Serial Decode ...
Страница 116: ...116 Agilent InfiniiVision 2000 X Series Oscilloscopes User s Guide 8 Display Settings ...
Страница 152: ...152 Agilent InfiniiVision 2000 X Series Oscilloscopes User s Guide 11 Trigger Mode Coupling ...
Страница 178: ...178 Agilent InfiniiVision 2000 X Series Oscilloscopes User s Guide 13 Cursors ...
Страница 214: ...214 Agilent InfiniiVision 2000 X Series Oscilloscopes User s Guide 16 Digital Voltmeter ...
Страница 226: ...226 Agilent InfiniiVision 2000 X Series Oscilloscopes User s Guide 17 Waveform Generator ...
Страница 238: ...238 Agilent InfiniiVision 2000 X Series Oscilloscopes User s Guide 18 Save Recall Setups Screens Data ...
Страница 244: ...244 Agilent InfiniiVision 2000 X Series Oscilloscopes User s Guide 19 Print Screens ...
Страница 280: ...280 Agilent InfiniiVision 2000 X Series Oscilloscopes User s Guide 21 Web Interface ...
Страница 334: ...334 Agilent InfiniiVision 2000 X Series Oscilloscopes User s Guide 24 I2C SPI Triggering and Serial Decode ...
Страница 352: ...352 Agilent InfiniiVision 2000 X Series Oscilloscopes User s Guide Index ...