316
Agilent InfiniiVision 2000 X-Series Oscilloscopes User's Guide
24
I2C/SPI Triggering and Serial Decode
5
For both the SCL (serial clock) and SDA (serial data) signals:
a
Connect an oscilloscope channel to the signal in the device under
test.
b
Press the
SCL
or
SDA
softkey; then, turn the Entry knob to select the
channel for the signal.
c
Press the corresponding
Threshold
softkey; then, turn the Entry knob
to select the signal threshold voltage level.
The threshold voltage level is used in decoding, and it will become
the trigger level when the trigger type is set to the selected serial
decode slot.
Data must be stable during the whole high clock cycle or it will be
interpreted as a start or stop condition (data transitioning while the
clock is high).
The SCL and SDA labels for the source channels are automatically set.
I2C Triggering
To set up the oscilloscope to capture I2C signals, see
"Setup for I2C
Signals"
on page 315.
After the oscilloscope has been set up to capture I2C signals, you can
trigger on a stop/start condition, a restart, a missing acknowledge, an
EEPROM data read, or on a read/write frame with a specific device
address and data value.
1
Press
[Trigger]
; then, select the
I2C
trigger type.
2
Press
[Trigger]
.
3
In the Trigger Menu, press the
Trigger
softkey; then, turn the Entry knob
to select the serial slot (Serial 1) on which the I
2
C signals are being
decoded.
Содержание 2000 X-Series
Страница 1: ...s1 Agilent InfiniiVision 2000 X Series Oscilloscopes User s Guide ...
Страница 20: ...20 Agilent InfiniiVision 2000 X Series Oscilloscopes User s Guide ...
Страница 44: ...44 Agilent InfiniiVision 2000 X Series Oscilloscopes User s Guide 1 Getting Started ...
Страница 58: ...58 Agilent InfiniiVision 2000 X Series Oscilloscopes User s Guide 2 Horizontal Controls ...
Страница 66: ...66 Agilent InfiniiVision 2000 X Series Oscilloscopes User s Guide 3 Vertical Controls ...
Страница 80: ...80 Agilent InfiniiVision 2000 X Series Oscilloscopes User s Guide 4 Math Waveforms ...
Страница 104: ...104 Agilent InfiniiVision 2000 X Series Oscilloscopes User s Guide 6 Digital Channels ...
Страница 110: ...110 Agilent InfiniiVision 2000 X Series Oscilloscopes User s Guide 7 Serial Decode ...
Страница 116: ...116 Agilent InfiniiVision 2000 X Series Oscilloscopes User s Guide 8 Display Settings ...
Страница 152: ...152 Agilent InfiniiVision 2000 X Series Oscilloscopes User s Guide 11 Trigger Mode Coupling ...
Страница 178: ...178 Agilent InfiniiVision 2000 X Series Oscilloscopes User s Guide 13 Cursors ...
Страница 214: ...214 Agilent InfiniiVision 2000 X Series Oscilloscopes User s Guide 16 Digital Voltmeter ...
Страница 226: ...226 Agilent InfiniiVision 2000 X Series Oscilloscopes User s Guide 17 Waveform Generator ...
Страница 238: ...238 Agilent InfiniiVision 2000 X Series Oscilloscopes User s Guide 18 Save Recall Setups Screens Data ...
Страница 244: ...244 Agilent InfiniiVision 2000 X Series Oscilloscopes User s Guide 19 Print Screens ...
Страница 280: ...280 Agilent InfiniiVision 2000 X Series Oscilloscopes User s Guide 21 Web Interface ...
Страница 334: ...334 Agilent InfiniiVision 2000 X Series Oscilloscopes User s Guide 24 I2C SPI Triggering and Serial Decode ...
Страница 352: ...352 Agilent InfiniiVision 2000 X Series Oscilloscopes User s Guide Index ...