Measurements
14
Agilent InfiniiVision 2000 X-Series Oscilloscopes User's Guide
191
Time Measurements
The following figure shows time measurement points.
The default lower, middle, and upper measurement thresholds are 10%,
50%, and 90% between Top and Base values. See
"Measurement
Thresholds"
on page 196 for other percentage threshold and absolute value
threshold settings.
•
"Period"
on page 191
•
"Frequency"
on page 192
•
"+ Width"
on page 193
•
"– Width"
on page 193
•
"Duty Cycle"
on page 193
•
"Rise Time"
on page 193
•
"Fall Time"
on page 193
•
"Delay"
on page 194
•
"Phase"
on page 195
Period
Period is the time period of the complete waveform cycle. The time is
measured between the middle threshold points of two consecutive,
like- polarity edges. A middle threshold crossing must also travel through
the lower and upper threshold levels which eliminates runt pulses. The X
cursors show what portion of the waveform is being measured. The Y
cursor shows the middle threshold point.
Rise Time
+ Width
Fall Time
- Width
Period
Middle
Upper
Thresholds
Lower
Содержание 2000 X-Series
Страница 1: ...s1 Agilent InfiniiVision 2000 X Series Oscilloscopes User s Guide ...
Страница 20: ...20 Agilent InfiniiVision 2000 X Series Oscilloscopes User s Guide ...
Страница 44: ...44 Agilent InfiniiVision 2000 X Series Oscilloscopes User s Guide 1 Getting Started ...
Страница 58: ...58 Agilent InfiniiVision 2000 X Series Oscilloscopes User s Guide 2 Horizontal Controls ...
Страница 66: ...66 Agilent InfiniiVision 2000 X Series Oscilloscopes User s Guide 3 Vertical Controls ...
Страница 80: ...80 Agilent InfiniiVision 2000 X Series Oscilloscopes User s Guide 4 Math Waveforms ...
Страница 104: ...104 Agilent InfiniiVision 2000 X Series Oscilloscopes User s Guide 6 Digital Channels ...
Страница 110: ...110 Agilent InfiniiVision 2000 X Series Oscilloscopes User s Guide 7 Serial Decode ...
Страница 116: ...116 Agilent InfiniiVision 2000 X Series Oscilloscopes User s Guide 8 Display Settings ...
Страница 152: ...152 Agilent InfiniiVision 2000 X Series Oscilloscopes User s Guide 11 Trigger Mode Coupling ...
Страница 178: ...178 Agilent InfiniiVision 2000 X Series Oscilloscopes User s Guide 13 Cursors ...
Страница 214: ...214 Agilent InfiniiVision 2000 X Series Oscilloscopes User s Guide 16 Digital Voltmeter ...
Страница 226: ...226 Agilent InfiniiVision 2000 X Series Oscilloscopes User s Guide 17 Waveform Generator ...
Страница 238: ...238 Agilent InfiniiVision 2000 X Series Oscilloscopes User s Guide 18 Save Recall Setups Screens Data ...
Страница 244: ...244 Agilent InfiniiVision 2000 X Series Oscilloscopes User s Guide 19 Print Screens ...
Страница 280: ...280 Agilent InfiniiVision 2000 X Series Oscilloscopes User s Guide 21 Web Interface ...
Страница 334: ...334 Agilent InfiniiVision 2000 X Series Oscilloscopes User s Guide 24 I2C SPI Triggering and Serial Decode ...
Страница 352: ...352 Agilent InfiniiVision 2000 X Series Oscilloscopes User s Guide Index ...