162
Agilent InfiniiVision 2000 X-Series Oscilloscopes User's Guide
12
Acquisition Control
Using Peak Detect Mode to Find a Glitch
1
Connect a signal to the oscilloscope and obtain a stable display.
2
To find the glitch, press the
[Acquire]
key; then, press the
Acq Mode
softkey until
Peak Detect
is selected.
3
Press the
[Display]
key then press the
∞
Persistence
(infinite persistence)
softkey.
Infinite persistence updates the display with new acquisitions but does
not erase previous acquisitions. New sample points are shown at normal
intensity while previous acquisitions are displayed at reduced intensity.
Waveform persistence is not kept beyond the display area boundary.
Press the
Clear Display
softkey to erase previously acquired points. The
display will accumulate points until
∞
Persistence
is turned off.
4
Characterize the glitch with Zoom mode:
a
Press the
zoom key (or press the
[Horiz]
key and then the
Zoom
softkey).
b
To obtain a better resolution of the glitch, expand the time base.
Figure 25
Sine With Glitch, Peak Detect Mode
Содержание 2000 X-Series
Страница 1: ...s1 Agilent InfiniiVision 2000 X Series Oscilloscopes User s Guide ...
Страница 20: ...20 Agilent InfiniiVision 2000 X Series Oscilloscopes User s Guide ...
Страница 44: ...44 Agilent InfiniiVision 2000 X Series Oscilloscopes User s Guide 1 Getting Started ...
Страница 58: ...58 Agilent InfiniiVision 2000 X Series Oscilloscopes User s Guide 2 Horizontal Controls ...
Страница 66: ...66 Agilent InfiniiVision 2000 X Series Oscilloscopes User s Guide 3 Vertical Controls ...
Страница 80: ...80 Agilent InfiniiVision 2000 X Series Oscilloscopes User s Guide 4 Math Waveforms ...
Страница 104: ...104 Agilent InfiniiVision 2000 X Series Oscilloscopes User s Guide 6 Digital Channels ...
Страница 110: ...110 Agilent InfiniiVision 2000 X Series Oscilloscopes User s Guide 7 Serial Decode ...
Страница 116: ...116 Agilent InfiniiVision 2000 X Series Oscilloscopes User s Guide 8 Display Settings ...
Страница 152: ...152 Agilent InfiniiVision 2000 X Series Oscilloscopes User s Guide 11 Trigger Mode Coupling ...
Страница 178: ...178 Agilent InfiniiVision 2000 X Series Oscilloscopes User s Guide 13 Cursors ...
Страница 214: ...214 Agilent InfiniiVision 2000 X Series Oscilloscopes User s Guide 16 Digital Voltmeter ...
Страница 226: ...226 Agilent InfiniiVision 2000 X Series Oscilloscopes User s Guide 17 Waveform Generator ...
Страница 238: ...238 Agilent InfiniiVision 2000 X Series Oscilloscopes User s Guide 18 Save Recall Setups Screens Data ...
Страница 244: ...244 Agilent InfiniiVision 2000 X Series Oscilloscopes User s Guide 19 Print Screens ...
Страница 280: ...280 Agilent InfiniiVision 2000 X Series Oscilloscopes User s Guide 21 Web Interface ...
Страница 334: ...334 Agilent InfiniiVision 2000 X Series Oscilloscopes User s Guide 24 I2C SPI Triggering and Serial Decode ...
Страница 352: ...352 Agilent InfiniiVision 2000 X Series Oscilloscopes User s Guide Index ...