Acquisition Control
12
Agilent InfiniiVision 2000 X-Series Oscilloscopes User's Guide
159
Memory Depth and Sample Rate
The number of points of oscilloscope memory is fixed, and there is a
maximum sample rate associated with oscilloscope's analog- to- digital
converter; however, the actual sample rate is determined by the time of
the acquisition (which is set according to the oscilloscope's horizontal
time/div scale).
sample rate = number of samples / time of acquisition
For example, when storing 50 µs of data in 50,000 points of memory, the
actual sample rate is 1 GSa/s.
Likewise, when storing 50 ms of data in 50,000 points of memory, the
actual sample rate is 1 MSa/s.
The actual sample rate is displayed in the right- side information area.
The oscilloscope achieves the actual sample rate by throwing away
(decimating) unneeded samples.
Selecting the Acquisition Mode
When selecting the oscilloscope acquisition mode, keep in mind that
samples are normally decimated at slower time/div settings.
At slower time/div settings, the effective sample rate drops (and the
effective sample period increases) because the acquisition time increases
and the oscilloscope's digitizer is sampling faster than is required to fill
memory.
For example, suppose an oscilloscope's digitizer has a sample period of
1 ns (maximum sample rate of 1 GSa/s) and a 1 M memory depth. At that
rate, memory is filled in 1 ms. If the acquisition time is 100 ms
(10 ms/div), only 1 of every 100 samples is needed to fill memory.
To select the acquisition mode:
1
Press the
[Acquire]
key on the front panel.
2
In the Acquire Menu, press the
Acq Mode
softkey; then, turn the Entry
knob to select the acquisition mode.
The InfiniiVision oscilloscopes have the following acquisition modes:
Содержание 2000 X-Series
Страница 1: ...s1 Agilent InfiniiVision 2000 X Series Oscilloscopes User s Guide ...
Страница 20: ...20 Agilent InfiniiVision 2000 X Series Oscilloscopes User s Guide ...
Страница 44: ...44 Agilent InfiniiVision 2000 X Series Oscilloscopes User s Guide 1 Getting Started ...
Страница 58: ...58 Agilent InfiniiVision 2000 X Series Oscilloscopes User s Guide 2 Horizontal Controls ...
Страница 66: ...66 Agilent InfiniiVision 2000 X Series Oscilloscopes User s Guide 3 Vertical Controls ...
Страница 80: ...80 Agilent InfiniiVision 2000 X Series Oscilloscopes User s Guide 4 Math Waveforms ...
Страница 104: ...104 Agilent InfiniiVision 2000 X Series Oscilloscopes User s Guide 6 Digital Channels ...
Страница 110: ...110 Agilent InfiniiVision 2000 X Series Oscilloscopes User s Guide 7 Serial Decode ...
Страница 116: ...116 Agilent InfiniiVision 2000 X Series Oscilloscopes User s Guide 8 Display Settings ...
Страница 152: ...152 Agilent InfiniiVision 2000 X Series Oscilloscopes User s Guide 11 Trigger Mode Coupling ...
Страница 178: ...178 Agilent InfiniiVision 2000 X Series Oscilloscopes User s Guide 13 Cursors ...
Страница 214: ...214 Agilent InfiniiVision 2000 X Series Oscilloscopes User s Guide 16 Digital Voltmeter ...
Страница 226: ...226 Agilent InfiniiVision 2000 X Series Oscilloscopes User s Guide 17 Waveform Generator ...
Страница 238: ...238 Agilent InfiniiVision 2000 X Series Oscilloscopes User s Guide 18 Save Recall Setups Screens Data ...
Страница 244: ...244 Agilent InfiniiVision 2000 X Series Oscilloscopes User s Guide 19 Print Screens ...
Страница 280: ...280 Agilent InfiniiVision 2000 X Series Oscilloscopes User s Guide 21 Web Interface ...
Страница 334: ...334 Agilent InfiniiVision 2000 X Series Oscilloscopes User s Guide 24 I2C SPI Triggering and Serial Decode ...
Страница 352: ...352 Agilent InfiniiVision 2000 X Series Oscilloscopes User s Guide Index ...