Cursors
13
Agilent InfiniiVision 2000 X-Series Oscilloscopes User's Guide
171
•
Track Waveform
— As you move a marker horizontally, the vertical
amplitude of the waveform is tracked and measured. The time and
voltage positions are shown for the markers. The vertical (Y) and
horizontal (X) differences between the markers are shown as
Δ
X and
Δ
Y values.
•
Binary
— Logic levels of displayed waveforms at the current X1 and
X2 cursor positions are displayed above the softkeys in binary. The
display is color coded to match the color of the related channel's
waveform.
•
Hex
— Logic levels of displayed waveforms at the current X1 and X2
cursor positions are displayed above the softkeys in hexadecimal.
Manual
and
Track Waveform
modes can be used on waveforms that are
displayed on the analog input channels (including math functions).
Binary
and
Hex
modes apply to digital signals (of MSO oscilloscope
models).
In
Hex
and
Binary
modes, a level can be displayed as 1 (higher than
trigger level), 0 (lower than trigger level), indeterminate state ( ), or X
(don't care).
In
Binary
mode, X is displayed if the channel is turned off.
In
Hex
mode, the channel is interpreted as 0 if turned off.
4
Press
Source
(or
X1 Source
,
X2 Source
in the
Track Waveform
mode); then,
select the input source for cursor values.
5
Select the cursor(s) to be adjusted:
•
Push the Cursors knob; then, turn the Cursors knob. To finalize your
selection, either push the Cursors knob again or wait about five
seconds for the popup menu to disappear.
Or:
Содержание 2000 X-Series
Страница 1: ...s1 Agilent InfiniiVision 2000 X Series Oscilloscopes User s Guide ...
Страница 20: ...20 Agilent InfiniiVision 2000 X Series Oscilloscopes User s Guide ...
Страница 44: ...44 Agilent InfiniiVision 2000 X Series Oscilloscopes User s Guide 1 Getting Started ...
Страница 58: ...58 Agilent InfiniiVision 2000 X Series Oscilloscopes User s Guide 2 Horizontal Controls ...
Страница 66: ...66 Agilent InfiniiVision 2000 X Series Oscilloscopes User s Guide 3 Vertical Controls ...
Страница 80: ...80 Agilent InfiniiVision 2000 X Series Oscilloscopes User s Guide 4 Math Waveforms ...
Страница 104: ...104 Agilent InfiniiVision 2000 X Series Oscilloscopes User s Guide 6 Digital Channels ...
Страница 110: ...110 Agilent InfiniiVision 2000 X Series Oscilloscopes User s Guide 7 Serial Decode ...
Страница 116: ...116 Agilent InfiniiVision 2000 X Series Oscilloscopes User s Guide 8 Display Settings ...
Страница 152: ...152 Agilent InfiniiVision 2000 X Series Oscilloscopes User s Guide 11 Trigger Mode Coupling ...
Страница 178: ...178 Agilent InfiniiVision 2000 X Series Oscilloscopes User s Guide 13 Cursors ...
Страница 214: ...214 Agilent InfiniiVision 2000 X Series Oscilloscopes User s Guide 16 Digital Voltmeter ...
Страница 226: ...226 Agilent InfiniiVision 2000 X Series Oscilloscopes User s Guide 17 Waveform Generator ...
Страница 238: ...238 Agilent InfiniiVision 2000 X Series Oscilloscopes User s Guide 18 Save Recall Setups Screens Data ...
Страница 244: ...244 Agilent InfiniiVision 2000 X Series Oscilloscopes User s Guide 19 Print Screens ...
Страница 280: ...280 Agilent InfiniiVision 2000 X Series Oscilloscopes User s Guide 21 Web Interface ...
Страница 334: ...334 Agilent InfiniiVision 2000 X Series Oscilloscopes User s Guide 24 I2C SPI Triggering and Serial Decode ...
Страница 352: ...352 Agilent InfiniiVision 2000 X Series Oscilloscopes User s Guide Index ...