![Agilent Technologies 2000 X-Series Скачать руководство пользователя страница 148](http://html.mh-extra.com/html/agilent-technologies/2000-x-series/2000-x-series_user-manual_2868575148.webp)
148
Agilent InfiniiVision 2000 X-Series Oscilloscopes User's Guide
11
Trigger Mode/Coupling
•
"To position the time reference (left, center, right)"
on page 54
To select the trigger coupling
1
Press the
[Mode/Coupling]
key.
2
In the Trigger Mode and Coupling Menu, press the
Coupling
softkey;
then, turn the Entry knob to select:
•
DC
coupling — allows DC and AC signals into the trigger path.
•
AC
coupling — places a 10 Hz high- pass filter in the trigger path
removing any DC offset voltage from the trigger waveform.
The high- pass filter in the External Trigger input path is 50 Hz for
all models.
Use AC coupling to get a stable edge trigger when your waveform has
a large DC offset.
•
LF
(low frequency)
Reject
coupling — adds a high- pass filter with the
3- dB point at 50 kHz in series with the trigger waveform.
Low frequency reject removes any unwanted low frequency
components from a trigger waveform, such as power line frequencies,
etc., that can interfere with proper triggering.
Use
LF Reject
coupling to get a stable edge trigger when your
waveform has low frequency noise.
•
TV
coupling — is normally grayed- out, but is automatically selected
when TV trigger is enabled in the Trigger Menu.
Pass
Band
50 kHz
0 dB
3 dB down point
DC
Содержание 2000 X-Series
Страница 1: ...s1 Agilent InfiniiVision 2000 X Series Oscilloscopes User s Guide ...
Страница 20: ...20 Agilent InfiniiVision 2000 X Series Oscilloscopes User s Guide ...
Страница 44: ...44 Agilent InfiniiVision 2000 X Series Oscilloscopes User s Guide 1 Getting Started ...
Страница 58: ...58 Agilent InfiniiVision 2000 X Series Oscilloscopes User s Guide 2 Horizontal Controls ...
Страница 66: ...66 Agilent InfiniiVision 2000 X Series Oscilloscopes User s Guide 3 Vertical Controls ...
Страница 80: ...80 Agilent InfiniiVision 2000 X Series Oscilloscopes User s Guide 4 Math Waveforms ...
Страница 104: ...104 Agilent InfiniiVision 2000 X Series Oscilloscopes User s Guide 6 Digital Channels ...
Страница 110: ...110 Agilent InfiniiVision 2000 X Series Oscilloscopes User s Guide 7 Serial Decode ...
Страница 116: ...116 Agilent InfiniiVision 2000 X Series Oscilloscopes User s Guide 8 Display Settings ...
Страница 152: ...152 Agilent InfiniiVision 2000 X Series Oscilloscopes User s Guide 11 Trigger Mode Coupling ...
Страница 178: ...178 Agilent InfiniiVision 2000 X Series Oscilloscopes User s Guide 13 Cursors ...
Страница 214: ...214 Agilent InfiniiVision 2000 X Series Oscilloscopes User s Guide 16 Digital Voltmeter ...
Страница 226: ...226 Agilent InfiniiVision 2000 X Series Oscilloscopes User s Guide 17 Waveform Generator ...
Страница 238: ...238 Agilent InfiniiVision 2000 X Series Oscilloscopes User s Guide 18 Save Recall Setups Screens Data ...
Страница 244: ...244 Agilent InfiniiVision 2000 X Series Oscilloscopes User s Guide 19 Print Screens ...
Страница 280: ...280 Agilent InfiniiVision 2000 X Series Oscilloscopes User s Guide 21 Web Interface ...
Страница 334: ...334 Agilent InfiniiVision 2000 X Series Oscilloscopes User s Guide 24 I2C SPI Triggering and Serial Decode ...
Страница 352: ...352 Agilent InfiniiVision 2000 X Series Oscilloscopes User s Guide Index ...