292
Agilent InfiniiVision 2000 X-Series Oscilloscopes User's Guide
22
Reference
Waveform Data
Header
A waveform may have more than one data set. Each waveform data set
will have a waveform data header. The waveform data header consists of
information about the waveform data set. This header is stored
immediately before the data set.
X Units
A 32-bit integer that identifies the unit of measure for X values in the acquired
data:
• 0 = Unknown.
• 1 = Volts.
• 2 = Seconds.
• 3 = Constant.
• 4 = Amps.
• 5 = dB.
• 6 = Hz.
Y Units
A 32-bit integer that identifies the unit of measure for Y values in the acquired
data. The possible values are listed above under X Units.
Date
A 16-byte character array, left blank in InfiniiVision oscilloscopes.
Time
A 16-byte character array, left blank in the InfiniiVision oscilloscopes.
Frame
A 24 byte character array that is the model number and serial number of the
oscilloscope in the format of: MODEL#:SERIAL#.
Waveform Label
A 16 byte character array that contains the label assigned to the waveform.
Time Tags
A 64-bit double, only used when saving multiple segments (requires
segmented memory option). This is the time (in seconds) since the first
trigger.
Segment Index
A 32-bit unsigned integer. This is the segment number. Only used when saving
multiple segments.
Waveform Data
Header Size
A 32-bit integer that is the size of the waveform data header.
Buffer Type
A 16-bit short that is the type of waveform data stored in the file:
• 0 = Unknown data.
• 1 = Normal 32-bit float data.
• 2 = Maximum float data.
• 3 = Minimum float data.
• 4 = Not used in InfiniiVision oscilloscopes.
• 5 = Not used in InfiniiVision oscilloscopes.
• 6 = Digital unsigned 8-bit char data (for digital channels).
Содержание 2000 X-Series
Страница 1: ...s1 Agilent InfiniiVision 2000 X Series Oscilloscopes User s Guide ...
Страница 20: ...20 Agilent InfiniiVision 2000 X Series Oscilloscopes User s Guide ...
Страница 44: ...44 Agilent InfiniiVision 2000 X Series Oscilloscopes User s Guide 1 Getting Started ...
Страница 58: ...58 Agilent InfiniiVision 2000 X Series Oscilloscopes User s Guide 2 Horizontal Controls ...
Страница 66: ...66 Agilent InfiniiVision 2000 X Series Oscilloscopes User s Guide 3 Vertical Controls ...
Страница 80: ...80 Agilent InfiniiVision 2000 X Series Oscilloscopes User s Guide 4 Math Waveforms ...
Страница 104: ...104 Agilent InfiniiVision 2000 X Series Oscilloscopes User s Guide 6 Digital Channels ...
Страница 110: ...110 Agilent InfiniiVision 2000 X Series Oscilloscopes User s Guide 7 Serial Decode ...
Страница 116: ...116 Agilent InfiniiVision 2000 X Series Oscilloscopes User s Guide 8 Display Settings ...
Страница 152: ...152 Agilent InfiniiVision 2000 X Series Oscilloscopes User s Guide 11 Trigger Mode Coupling ...
Страница 178: ...178 Agilent InfiniiVision 2000 X Series Oscilloscopes User s Guide 13 Cursors ...
Страница 214: ...214 Agilent InfiniiVision 2000 X Series Oscilloscopes User s Guide 16 Digital Voltmeter ...
Страница 226: ...226 Agilent InfiniiVision 2000 X Series Oscilloscopes User s Guide 17 Waveform Generator ...
Страница 238: ...238 Agilent InfiniiVision 2000 X Series Oscilloscopes User s Guide 18 Save Recall Setups Screens Data ...
Страница 244: ...244 Agilent InfiniiVision 2000 X Series Oscilloscopes User s Guide 19 Print Screens ...
Страница 280: ...280 Agilent InfiniiVision 2000 X Series Oscilloscopes User s Guide 21 Web Interface ...
Страница 334: ...334 Agilent InfiniiVision 2000 X Series Oscilloscopes User s Guide 24 I2C SPI Triggering and Serial Decode ...
Страница 352: ...352 Agilent InfiniiVision 2000 X Series Oscilloscopes User s Guide Index ...