Agilent InfiniiVision 2000 X-Series Oscilloscopes User's Guide
1
Getting Started
16.
Vertical
controls
The Vertical controls consist of:
• Analog channel on/off keys — Use these keys to switch a channel on
or off, or to access a channel's menu in the softkeys. There is one
channel on/off key for each analog channel.
• Vertical scale knob — There are knobs marked
for each
channel. Use these knobs to change the vertical sensitivity (gain) of
each analog channel.
• Vertical position knobs — Use these knobs to change a channel's
vertical position on the display. There is one Vertical Position control
for each analog channel.
• [Label] key — Press this key to access the Label Menu, which lets
you enter labels to identify each trace on the oscilloscope display. See
Chapter 9
, “Labels,” starting on page 117.
For more information, see
Chapter 3
, “Vertical Controls,” starting on
page 59.
17.
Analog channel
inputs
Attach oscilloscope probes or BNC cables to these BNC connectors.
In the InfiniiVision 2000 X-Series oscilloscopes, the analog channel
inputs have 1 M
Ω
impedance.
Also, there is no automatic probe detection, so you must properly set the
probe attenuation for accurate measurement results. See
“To specify the
probe attenuation"
on page 64.
18.
Demo 2,
Ground, and
Demo 1
terminals
• Demo 2 terminal — This terminal outputs the Probe Comp signal
which helps you match a probe's input capacitance to the
oscilloscope channel to which it is connected. See
“Compensate
Passive Probes"
on page 29. With certain licensed features, the
oscilloscope can also output demo or training signals on this terminal.
• Ground terminal — Use the ground terminal for oscilloscope probes
connected to the Demo 1 or Demo 2 terminals.
• Demo 1 terminal — With certain licensed features, the oscilloscope
can output demo or training signals on this terminal.
Содержание 2000 X-Series
Страница 1: ...s1 Agilent InfiniiVision 2000 X Series Oscilloscopes User s Guide ...
Страница 20: ...20 Agilent InfiniiVision 2000 X Series Oscilloscopes User s Guide ...
Страница 44: ...44 Agilent InfiniiVision 2000 X Series Oscilloscopes User s Guide 1 Getting Started ...
Страница 58: ...58 Agilent InfiniiVision 2000 X Series Oscilloscopes User s Guide 2 Horizontal Controls ...
Страница 66: ...66 Agilent InfiniiVision 2000 X Series Oscilloscopes User s Guide 3 Vertical Controls ...
Страница 80: ...80 Agilent InfiniiVision 2000 X Series Oscilloscopes User s Guide 4 Math Waveforms ...
Страница 104: ...104 Agilent InfiniiVision 2000 X Series Oscilloscopes User s Guide 6 Digital Channels ...
Страница 110: ...110 Agilent InfiniiVision 2000 X Series Oscilloscopes User s Guide 7 Serial Decode ...
Страница 116: ...116 Agilent InfiniiVision 2000 X Series Oscilloscopes User s Guide 8 Display Settings ...
Страница 152: ...152 Agilent InfiniiVision 2000 X Series Oscilloscopes User s Guide 11 Trigger Mode Coupling ...
Страница 178: ...178 Agilent InfiniiVision 2000 X Series Oscilloscopes User s Guide 13 Cursors ...
Страница 214: ...214 Agilent InfiniiVision 2000 X Series Oscilloscopes User s Guide 16 Digital Voltmeter ...
Страница 226: ...226 Agilent InfiniiVision 2000 X Series Oscilloscopes User s Guide 17 Waveform Generator ...
Страница 238: ...238 Agilent InfiniiVision 2000 X Series Oscilloscopes User s Guide 18 Save Recall Setups Screens Data ...
Страница 244: ...244 Agilent InfiniiVision 2000 X Series Oscilloscopes User s Guide 19 Print Screens ...
Страница 280: ...280 Agilent InfiniiVision 2000 X Series Oscilloscopes User s Guide 21 Web Interface ...
Страница 334: ...334 Agilent InfiniiVision 2000 X Series Oscilloscopes User s Guide 24 I2C SPI Triggering and Serial Decode ...
Страница 352: ...352 Agilent InfiniiVision 2000 X Series Oscilloscopes User s Guide Index ...