99
Automatic sensitivity adjustment
Display settings: MENU key
Sensitivity adjustment button
Automatic sensitivity adjustment button
[Automatic sensitivity adjustment] settings
screen
1
2
3
Name
Setting value (setting range)
Description
1
Automatic sensitivity
adjustment execution
condition setting field
OFF, Temperature change ,
Preset time 1,
Preset time 1&2,
Interval
Set the automatic sensitivity adjustment
execution condition.
If you select the preset time or interval, the
relevant time setting is displayed below the
setting field.
2
Automatic sensitivity
adjustment time setting 0:00 - 24:00
Set the automatic sensitivity adjustment
start time.
3
Automatic sensitivity
adjustment interval
setting
0.5 ,
0
1.0,
0
1.5,
0
2.0,
0
2.5,
0
3.0,
0
3.5,
0
4.0,
0
4.5,
0
5.0,
0
5.5,
0
6.0,
0
7.0,
0
8.0,
0
9.0,
10.0, 11.0, 12.0, 14.0, 16.0,
18.0, 20.0, 22.0, 24.0
Set the automatic sensitivity adjustment
interval.
Settings in the red box are default values (factory settings).
This function automatically adjusts the sensitivity of the balance according to ambient temperature
change, set time or interval time using the internal weight. It works even when the display is off. If the
GLP output is enabled, the sensitivity adjustment execution record is output after the sensitivity
adjustment.
The execution condition for the automatic sensitivity adjustment can be based on the temperature
change, preset time or interval.
For preset time, you can configure two settings: “Preset time 1” and “Preset time 2”.
For the interval, you can set a value between 0.5h and 24.0h.
Caution
If something is on the weighing pan, the balance itself will judge that it is in use and will not perform
automatic sensitivity adjustment.
The criteria for performing automatic sensitivity adjustment are as follows.
BA-6TE / BA-6DTE
Less than 0.1 g
BA-225DTE / BA-125DTE / BA-225TE
To maintain the correct sensitivity adjustment of the balance, do not place anything on the weighing pan
while not in use.
Содержание BA-T Series
Страница 1: ...BA 6TE BA 6DTE BA 225TE BA 225DTE BA 125DTE 1WMPD4004315C...
Страница 5: ...5 11 Calibration test check 101 12 Filter Settings 135 13 System settings 137...
Страница 107: ...107 Output example for daily check output results Daily check output results Daily check PDF results...
Страница 117: ...117 Output example for periodic check output results Periodic check output results...
Страница 118: ...118 Periodic check results PDF...
Страница 129: ...129 AND MEET analysis graph example...
Страница 215: ...215 Unit mm Display unit 138 182 73 Ionizer 162 68 129...
Страница 219: ......