41
Percent mode sample input [Sample input mode] screen
Display settings: HOME key
APP button
[Application] settings screen
Application selection button Select [Percent mode]
HOME key
Percent mode screen Percent mode setting button
Sample input button [Sample input mode] screen
1
Sample input mode display
2
SAVE button
3
Back button
Name
Description
1
Sample input mode
display
Displayed in the Percent mode sample input mode.
2
SAVE button
Stores the 100% reference mass.
3
Back button
Displays the [Percent mode setting] screen.
This screen is used to input the 100% reference mass for the Percent mode by weighing samples.
How to store the reference mass
Weighing example for BA-6DTE
Step 1. Place a container on the weighing pan if
necessary and press the RE-ZERO
button
to set the weighing value to
zero.
Step 2. Place the specified number of sample pieces
on the container/weighing pan.
Step 3. Press the SAVE button
. “The 100 %
reference mass has been stored.” is
displayed.
Press the HOME key
to return to the
Percent mode screen.
Weighing example
When samples are weighed,
the percent is displayed.
Container (tare)
Weighing pan
RE-ZERO button
Zero display
Weighing 100% reference mass samples
SAVE button
Storage complete
Percent display
Caution
If the balance judges that the sample reference mass for 100% is too light to be stored, the SAVE
button
is disabled.
The decimal separator position varies according to the 100% reference mass.
The stored 100% reference mass is stored in nonvolatile memory even if the power is removed.
Содержание BA-T Series
Страница 1: ...BA 6TE BA 6DTE BA 225TE BA 225DTE BA 125DTE 1WMPD4004315C...
Страница 5: ...5 11 Calibration test check 101 12 Filter Settings 135 13 System settings 137...
Страница 107: ...107 Output example for daily check output results Daily check output results Daily check PDF results...
Страница 117: ...117 Output example for periodic check output results Periodic check output results...
Страница 118: ...118 Periodic check results PDF...
Страница 129: ...129 AND MEET analysis graph example...
Страница 215: ...215 Unit mm Display unit 138 182 73 Ionizer 162 68 129...
Страница 219: ......