18. Connecting to a Personal Computer
Quick USB mode
The quick USB mode is a function where a balance connected to a personal computer with a USB cable
and inputs the output of the balance directly to the software on the personal computer, such as Excel or
Word. The supported OS is Windows 7 or later.
Since the Windows standard driver (HID) is used, there is no need to install a dedicated driver and
communication is possible simply by connecting.
Caution
Quick USB is a one-way communication from the balance to the personal computer. It is no possible to
send commands to control the balance from the personal computer.
Turn off the screen saver and standby mode of the personal computer.
Do not use Quick USB when the data output mode of the balance is set to the stream mode.
In stream mode, the balance keeps outputting weighing data to the personal computer. It may cause
unintended operation on the computer.
Setting procedure
To use the Quick USB mode, set the function table of the balance to [Quick USB mode].
MENU key
System settings button
Communication button
USB interface button
Select in the [USB interface] screen.
USB output format
When USB is used, select the output format in the [
] screen.
For details on the output format, refer to "
” (function table).
Setting procedure (to send weighing data with the PRINT button
of the balance)
Step 1. In the function table of the balance, set to [Quick USB].
Step 2. Connect the balance to the personal computer with the included USB cable.
Step 3. When the personal computer is connected to the balance for the first time, the computer will
automatically start installing the driver.
Step 4. Start the software (Excel, etc.) used for weighing data transmission on the computer.
Step 5. Be sure to set the keyboard to single-byte input mode. (Data cannot be entered correctly in the
double-byte input mode.)
Step 6. Place the cursor where you want to enter the weighing data.
Step 7. Press the PRINT button
of the balance to send weighing data from the balance and enter it
at the cursor position.
Step 8. To finish data transmission, disconnect the USB cable.
Содержание BA-T Series
Страница 1: ...BA 6TE BA 6DTE BA 225TE BA 225DTE BA 125DTE 1WMPD4004315C...
Страница 5: ...5 11 Calibration test check 101 12 Filter Settings 135 13 System settings 137...
Страница 107: ...107 Output example for daily check output results Daily check output results Daily check PDF results...
Страница 117: ...117 Output example for periodic check output results Periodic check output results...
Страница 118: ...118 Periodic check results PDF...
Страница 129: ...129 AND MEET analysis graph example...
Страница 215: ...215 Unit mm Display unit 138 182 73 Ionizer 162 68 129...
Страница 219: ......