121
Internal calibration test
Display settings: MENU key
Calibration test/check button
Internal calibration test button
[Calibration test] screen
2
3
1
4
Name
Description
1
Back button
Returns to the previous screen.
2
Calibration test result display
Displays the measurement result.
3
PRINT button
Outputs the measurement result to the device connected to the
balance.
The result is automatically output if [GLP output] in [Data output]
is set to ON.
4
Exit button
Exits the screen that shows the measurement result.
The internal weight is used to perform the calibration test.
Be sure to warm up the balance with nothing on the weighing pan for at least an hour, or at least
four hours for BA-6TE/BA-6DTE, with the AC adapter connected to the power supply.
Do not apply vibration and the like to the balance during the calibration test.
When the test is completed, the [Calibration test result] screen is displayed.
Caution
No sensitivity adjustment is performed.