36
Counting mode sample input
[Sample input mode] screen
Display settings: HOME key
APP button
[Application] settings screen
Application selection button Select [Counting mode].
HOME key
[Counting mode] screen Unit weight setting button
Input by weighing samples button Counting mode [Sample input mode] screen
1
Sample input mode display
2
Number of samples input button
3
Back button
4
SAVE button
Name
Description
1
Sample input mode
display
Displayed in the Counting mode sample input mode.
2
Number of samples
input button
Input the number of samples for the input by weighing samples.
Input range: 10 - 10000 pcs
(10 - 100 pcs with the touch screen software version 1.020 or earlier)
3
Back button
Displays the [Counting mode] screen.
4
SAVE button
Store the unit weight based on the current weight value and number of samples.
This screen is used to input the unit weight by weighing samples for the Counting mode.
How to store the unit weight
Counting example for BA-6DTE
Step 1. Press the Number of samples input
button.
*
Note that a greater number of sample
pieces will yield more accurate counting
result since the sample unit weight is
usually considered to vary more or less.
Step 2. Place a container on the weighing pan
and press the RE-ZERO button
to
set the weighing value to zero.
Step 3. Place the specified number of sample
pieces on the container/weighing pan.
Step 4. Press the SAVE button
. When “The
unit weight has been stored.” is displayed,
the storage is complete.
Press the HOME key
to return to the
Counting mode screen.
Number of samples input button
Numerical value input screen for
inputting the number of samples
Container (tare)
RE-ZERO button
Zero display
Weighing with the specified number of samples
SAVE button
Storage complete
Counting mode display
Caution
If the balance judges that the unit weight is too light to be stored, the SAVE button
is disabled.
The stored unit weight is stored in nonvolatile memory even if the power is removed.
Содержание BA-T Series
Страница 1: ...BA 6TE BA 6DTE BA 225TE BA 225DTE BA 125DTE 1WMPD4004315C...
Страница 5: ...5 11 Calibration test check 101 12 Filter Settings 135 13 System settings 137...
Страница 107: ...107 Output example for daily check output results Daily check output results Daily check PDF results...
Страница 117: ...117 Output example for periodic check output results Periodic check output results...
Страница 118: ...118 Periodic check results PDF...
Страница 129: ...129 AND MEET analysis graph example...
Страница 215: ...215 Unit mm Display unit 138 182 73 Ionizer 162 68 129...