110
4. Weighing pan check
Make sure that there is nothing on the weighing pan.
1
Skip button
2
Back button
3
Confirm button
Name
Description
1
Skip button
Skips this check item and proceeds to the next item.
2
Back button
Returns to the previous screen.
3
Confirm button
Considers this check item as OK and proceeds to the next item.
5. Calibration Test (Before Sensitivity Adjustment)
The calibration test is performed before sensitivity adjustment.
When the test is completed, the [Calibration test result] screen is displayed.
When the internal weight is used, this is automatically performed. When an external weight is used,
follow the instruction.
Do not apply vibration and the like to the balance during the calibration test.
3
1
4
2
5
6
Name
Setting value
(setting range)
Description
1
Weight to be
used
Internal weight ,
External weight
Select the weight to be used.
2
Back button
-
Cancels and returns to the previous screen.
3
Result display
-
Displays the calibration test result.
4
Skip button
-
Skips this check item and proceeds to the next item.
5
Execute button
-
Executes the calibration test.
6
PRINT button
Outputs the results to the device connected to the balance.
The results are automatically output if [GLP output] in [Data
output] is set to ON.
Settings in the red box are default values (factory settings).
Содержание BA-T Series
Страница 1: ...BA 6TE BA 6DTE BA 225TE BA 225DTE BA 125DTE 1WMPD4004315C...
Страница 5: ...5 11 Calibration test check 101 12 Filter Settings 135 13 System settings 137...
Страница 107: ...107 Output example for daily check output results Daily check output results Daily check PDF results...
Страница 117: ...117 Output example for periodic check output results Periodic check output results...
Страница 118: ...118 Periodic check results PDF...
Страница 129: ...129 AND MEET analysis graph example...
Страница 215: ...215 Unit mm Display unit 138 182 73 Ionizer 162 68 129...