45
Manual (external weight)
For this setting, measurement is performed with your weight to compute the minimum weight.
The procedure is displayed in the message field.
Weighing example for BA-6DTE
Step 1. Press the RE-ZERO button
.
Step 2. Place a weight on the weighing pan.
Step 3. Remove the weight when the stabilization
indicator is displayed.
Step 4. Repeat steps 1, 2 and 3 ten times.
Step 5. When the measurement is complete, the
[Minimum weight measurement result] screen
described in “
” is displayed.
Weighing pan
RE-ZERO button
Place the weight
Remove
the weight
10 times complete
[Minimum weight measurement result] screen