46
Minimum weight measurement result
Display settings: HOME key
APP button
[Application] settings screen
Application selection button Select [Minimum weight alert function]
HOME key
Weighing screen with the minimum weight alert function
Minimum weight setting button
Input by measuring repeatability button [Minimum weight setting] screen
Measure button [Minimum weight measurement result] screen.
The [Minimum weight measurement result] screen is displayed when the
1
Measurement method
2
Measurement tolerance
3
Repeatability (standard deviation)
4
Minimum weight
5
PRINT button
6
Register button
Name
Description
1
Measurement method Displays the method used for the minimum weight measurement.
2
Measurement tolerance Displays the minimum weight measurement tolerance.
3
Repeatability (standard
deviation)
Displays the standard deviation for this measurement.
4
Minimum weight
Displays the minimum weight.
5
PRINT button
Outputs the minimum weight measurement result to the device connected to
the balance.
6
Register button
Registers the minimum weight.
*
If the repeatability (standard deviation) is 0.41d or less, the minimum weight is 2000 times the 0.41d in
compliance with USP.
"d" is a unit of readability.
Содержание BA-T Series
Страница 1: ...BA 6TE BA 6DTE BA 225TE BA 225DTE BA 125DTE 1WMPD4004315C...
Страница 5: ...5 11 Calibration test check 101 12 Filter Settings 135 13 System settings 137...
Страница 107: ...107 Output example for daily check output results Daily check output results Daily check PDF results...
Страница 117: ...117 Output example for periodic check output results Periodic check output results...
Страница 118: ...118 Periodic check results PDF...
Страница 129: ...129 AND MEET analysis graph example...
Страница 215: ...215 Unit mm Display unit 138 182 73 Ionizer 162 68 129...
Страница 219: ......